Calibration

74

6. Press

ENTER

to end the calibration. The display will show the

measured value alternating with Hold.

7. Press

ENTER

to return to the measuring mode. The outputs will

remain in the hold mode for approximately 20 seconds.

Note: For comparison calibration via Foundation Fieldbus, see

section 8.1.1.4 on page 73

.

8.1.1.5

Zero adjustment for ISFET probes—For si792(x) P (HART)
and si792x P-FF only

The nominal zero point must be adjusted each time a new ISFET or
probe with an offset or isopotential other than pH 7 is connected. If not
adjusted, the Sensoface messages will not be accurate. Complete the
following steps to adjust the zero point.

1. Press

CAL

, enter passcode: 1001,

ENTER

. The display will show

CAL ZRO and then the pH value for the zero point.

2. Place the pH probe in a pH 7.00 buffer solution. Briefly stir the

solution with the probe.

3. Enter the temperature-corrected pH value of the buffer solution. The

value must be between 6.50 and 7.50 pH. Press

ENTER

to save

the value.

4. The transmitter will measure the mV reading of the probe. The

hourglass will flash. When complete, the mV value of the zero offset
of the probe will be displayed. Press

ENTER

.

5. Press

ENTER

to end the zero adjustment.

After adjusting the zero offset, be sure to calibrate the probe by
following one of the available procedures:

Automatic buffer recognition

Manual buffer entry

Product comparison method

Data entry