2 orp calibration, Calibration, Mv δ – Hach-Lange SI792 P, SI792x P, SI792x P-FF, SI792x P-PA User Manual User Manual

Page 77

Advertising
background image

Calibration

75

8.1.2

ORP calibration

The ORP probe is calibrated using an ORP (redox) standard solution.
The probe can also be calibrated by product comparison
(

section 8.1.1.4 on page 73

).

During calibration, the difference between the measured potential and
the specified potential of the calibration standard is determined. In the
measurement mode, this difference is added to the measured potential
to obtain the displayed potential (equation

(1)

).

(1)

where:
mV

ORP

= displayed ORP

mV

meas

= direct probe potential

ΔmV = delta value, determined during calibration

The probe potential can also be related to another reference system
such as the standard hydrogen probe. In this case the temperature-
corrected potential (

Table 20

) of the reference probe used must be

entered during calibration. During measurement, this value is then
added to the measured potential.

Table 20 Temperature-corrected potential of

commonly used reference systems

Temperature

(°C)

Ag/AgCl/KCl

1 mol/L

(

Δ

mV)

Ag/AgCl/KCl

3 mol/L

(

Δ

mV)

Thalamid

(

Δ

mV)

Mercuric

sulfate
(

Δ

mV)

0

249

224

–559

672

10

244

217

–564

664

20

240

211

–569

655

25

236

207

–571

651

30

233

203

–574

647

40

227

196

–580

639

50

221

188

–585

631

60

214

180

–592

623

mV

ORP

mV

meas

mV

Δ

+

=