1 output range, 2 time averaging filter, 1 output range 5.2.1.2 time averaging filter – Hach-Lange SI792 P, SI792x P, SI792x P-FF, SI792x P-PA User Manual User Manual

Page 39: Section 5.2.1.1, Section 5.2.1.2, Operation—4–20 ma/hart

Advertising
background image

Operation—4–20 mA/HART

37

Press

ENTER

to access a menu item. Use the

ARROW KEYS

to edit

values. Press

ENTER

to save the settings. If a value is outside of the

acceptable range, “Err” will be displayed and the value will not be
accepted. To exit the menu and return to the measurement mode, press

CONF

and then

ENTER

.

Example: Set the 4 mA current output signal to 3 pH.

1. Press

CONF

, enter passcode: 1200,

ENTER

. The display will show

Conf and then out.1MNU.

2. Press

ENTER

to access the output setup menu. The display will

show o1.UniT.

3. Press

ENTER

,

ENTER

to show the 4 mA submenu. The display will

show o1.4mA.

4. Use the

UP ARROW

and

RIGHT ARROW

to edit the value to

03.00 pH. Press

ENTER

to save the value. The display will show

o1.20mA.

5. Press

CONF

and then

ENTER

to exit the configuration menu.

5.2.1.1

Output range

The upper and lower end of the pH or ORP measurement range should
correspond to the 4 mA and 20 mA signals. For example, to set a range
of 3–12 pH, set the 4 mA signal to correspond to a pH value of 3 and
the 20 mA signal to correspond to a pH value of 12.

5.2.1.2

Time averaging filter

An averaging filter is available to reduce noise in the output signal. The
filter averages readings over a specified time interval. The time interval
can be set from 0 to 120 seconds (default: 0 seconds).

When set to 0 seconds, there is no signal averaging for noise reduction.
When set to 120 seconds, the current output value will be correspond to
the process signal averaged over the last 120 seconds. Increase the
time interval to reduce the noise in the output signal.

Note: The filter acts on the output signal but not on the displayed value.