Troubleshooting

86

10.5 Diagnostic functions

Table 25

lists the diagnostic functions.

Table 25 Diagnostic functions

Function

Description

View the output
current

From the measuring mode, press

ENTER

. The output

current will show in the main display for five seconds, then
return to measuring mode.

View calibration
data

From the measuring mode, press

CAL

and enter passcode

‘0000’. The slope will show in the main display and the zero
current will show in the secondary display. After 20 seconds,
the transmitter will return to the measuring mode, or press

ENTER

to return immediately to the measuring mode.

View the last error
message

From the measuring mode, press

CONF

and enter passcode

‘0000’. The last error message will show for 20 seconds, or
press

ENTER

to return immediately to the measuring mode.

After viewing the error, the message will be deleted.

View the
uncorrected probe
potential (sensor
monitor)

From the measuring mode, press

CONF

and enter passcode

‘2222’. The measured (uncorrected) potential will show in
the main display. Press

ENTER

to return to the

measurement mode.

Specify output
current

From the measuring mode, press

CONF

and enter passcode

‘5555’. The measured current will show in the secondary
display. The output current in the main display can be
modified.

To change the output current:

1. Select the current value with the

RIGHT ARROW

key.

2. Edit the number using the

UP ARROW

key.

3. Press

ENTER

to confirm. The entered value will show in

the secondary display. The transmitter will remain in Hold
mode.

4. To exit Hold mode, press

CONF

, then

ENTER

to return to

measuring mode.