Operation—Foundation Fieldbus

47

6.2.2

Standard analog input block (AI)

Three Analog Input Function Blocks provide for cyclic transmission of
measured values (currently measured value with status, alarm limits,
freely selectable process parameter).

6.2.2.1 Operating modes

Use the MODE_BLK parameter to set the following operating modes:

OOS—out of service. If not write-protected, access to all
parameters is allowed.

MAN—manual

Auto—online, normal state

6.2.2.2 Set the parameter and units

Use CHANNEL to set the measured parameter and units (

Table 12

).

The corresponding measurement unit is selected in the UNITS
subparameter of XD_SCALE (

Table 13 on page 49

).

Table 12 Measurement parameters and units

Channel

Parameter

Unit

1

pH

pH

2

ORP

mV

3

Temperature

°C

°F

4

Glass impedance

M

Ω

5

Ref. impedance

k

Ω

6

Slope

%

7

Asymmetry potential

mV