Installation
26
3.5.3.1
Hach combination pH/ORP sensors wiring
1. Install the jumpers shown in
Figure 6
.
2. Use
to wire the sensor to the transmitter.
Figure 6 Wiring for Hach combination pH/ORP sensors
1 Jumper between terminal 4 and 5
3 Cable shield connection
Note: Do not connect to earth ground.
Table 3 Terminal assignments—pH/ORP sensors
Terminal
No.
Assignment
Hach PC and RC
series color
Six-plug
connection
1
Measure electrode
coaxial cable center
wire
coaxial cable center
wire
4
Reference electrode
braided coaxial shield
braided coaxial
shield