1 hach combination ph/orp sensors wiring, Ensors, Installation – Hach-Lange SI792 P, SI792x P, SI792x P-FF, SI792x P-PA User Manual User Manual

Page 28

Advertising
background image

Installation

26

3.5.3.1

Hach combination pH/ORP sensors wiring

1. Install the jumpers shown in

Figure 6

.

2. Use

Table 3

to wire the sensor to the transmitter.

Figure 6 Wiring for Hach combination pH/ORP sensors

1 Jumper between terminal 4 and 5

3 Cable shield connection

Note: Do not connect to earth ground.

2 Wiring terminals—see

Table 3

Table 3 Terminal assignments—pH/ORP sensors

Terminal

No.

Assignment

Hach PC and RC

series color

Six-plug

connection

1

Measure electrode

coaxial cable center

wire

coaxial cable center

wire

4

Reference electrode

braided coaxial shield

braided coaxial

shield

Advertising