Operation—Profibus PA
64
7.1.2
Configuration menu
Select process variable pH/mV
pH (default) Range: 0.00–14.00 A change of the process variable
requires a complete configura-
tion
mV Range: –1500–1500
Select the temperature unit
Auto
Auto °C (default)
The measurement and
calibration are automatically
recorded when the temperature
sensor is connected
Auto °F
man
man °C
Manual input during
measurement and calibration
man °F
manAuto
°C Auto man
Recorded automatically during
measurement
Manual input during calibration
°F Auto man
Select the temperature sensor (auto)
PT100
PT1000 (default for EU version)
300 NTC (default for standard version with pHD and LCP sensors)
BUS EXT
External temperature during
measurement (°C)
Manual input during calibration
(°C)
Enter temperature of sensor (man)
xxx.x °C (default: 025.0 °C)
xxx.x °F
Select Sensocheck
CHECK ON
Continuous Sensocheck
evaluation of sensor function
CHECK OFF (default)