Select process variable pH/mV

pH (default) Range: 0.00–14.00 A change of the process variable

requires a complete configura-
tion

mV Range: –1500–1500

Select the temperature unit

Auto

Auto °C (default)

The measurement and
calibration are automatically
recorded when the temperature
sensor is connected

Auto °F

man

man °C

Manual input during
measurement and calibration

man °F

manAuto

°C Auto man

Recorded automatically during
measurement

Manual input during calibration

°F Auto man

Select the temperature sensor (auto)

PT100

PT1000 (default for EU version)

300 NTC (default for standard version with pHD and LCP sensors)

BUS EXT

External temperature during
measurement (°C)

Manual input during calibration
(°C)

Enter temperature of sensor (man)

xxx.x °C (default: 025.0 °C)

xxx.x °F

Select Sensocheck

CHECK ON

Continuous Sensocheck
evaluation of sensor function

CHECK OFF (default)