3 output signal during errors, 4 output signal during hold, Section 5.2.1.3 – Hach-Lange SI792 P, SI792x P, SI792x P-FF, SI792x P-PA User Manual User Manual
Page 40: Section 5.2.1.4, Operation—4–20 ma/hart
Advertising
Operation—4–20 mA/HART
38
5.2.1.3
Output signal during errors
When an error condition occurs, a 22 mA output signal can be sent as
a notification (default: off).
5.2.1.4
Output signal during HOLD
The output signal during hold periods can be maintained at the last
measured value (
) or fixed at a specified value (
).
The allowable range for the fixed value is 3.4 to 22 mA.
Figure 12 Output signal during HOLD—last value
1 Output signal during HOLD
2 HOLD mode time span
Figure 13 Output signal during HOLD—fixed value
1 Output signal during HOLD
2 HOLD mode time span