Operation—4–20 mA/HART

38

5.2.1.3

Output signal during errors

When an error condition occurs, a 22 mA output signal can be sent as
a notification (default: off).

5.2.1.4

Output signal during HOLD

The output signal during hold periods can be maintained at the last
measured value (

Figure 12

) or fixed at a specified value (

Figure 13

).

The allowable range for the fixed value is 3.4 to 22 mA.

Figure 12 Output signal during HOLD—last value

1 Output signal during HOLD

2 HOLD mode time span

Figure 13 Output signal during HOLD—fixed value

1 Output signal during HOLD

2 HOLD mode time span