Calibration

76

Be sure that ORP is selected in the Configure, Output menu
(

section 5.2.1 on page 36

).

Be sure that the measurement and calibration temperature are the
same. The temperature behavior of the reference probe is not
automatically taken into account.

Complete the following steps to calibrate the ORP probe:

1. Press

CAL

, enter passcode: 1100,

ENTER

. The display will show the

mV screen to be edited.

2. Remove the ORP and temperature probes from the process. Clean

the probes and immerse in the redox standard. Briefly stir the
solution with the probes.

3. Enter the millivolt value of the redox standard solution.

4. Press

ENTER

. The display will show the delta mV value.

5. Press

ENTER

. The display will show the measured mV value

alternating with Hold.

6. Press

ENTER

to return to the measurement mode.

7. Remove the probes from the solution, rinse and return to the

process. The HOLD mode will remain active for 20 seconds.