Edit lot, Performing system maintenance, Performing individual maintenance commands – Luminex xPONENT 3.1 Rev 2 User Manual

Page 131: Perform system calibration, Manage maintenance routines, Create new routine

Advertising
background image

Edit Lot

1. Open the Protocols page, then open the Protocols tab. Click the Stds & Ctrls tab.

2. In the Installed Kits And Lots section, select a lot and then click Edit. The Std/Ctrl

Details tab opens. Change the lot information as appropriate.

Performing System Maintenance

Performing Individual Maintenance Commands

Open the Maintenance page, then click the Auto Maint tab to display the automated
maintenance options. To perform an individual maintenance command, click on the
appropriate button in the Cmds & Routines tab. For a description of each of these
commands, see

Cmds & Routines Tab

WARNING:

Lack of proper maintenance can lead to inaccurate test results
and failure of system components. Please follow all scheduled
maintenance routines to ensure proper functioning of
equipment.

Perform System Calibration

Perform system calibration weekly. In addition, recalibrate the system if the delta cal
temperature exceeds +/-3ºC. For further information, see

System Initialization

.

Manage Maintenance Routines

You manage maintenance routines at the Cmds & Routines tab of the Maintenance page.
Open the Maintenance page, then click the Cmds & Routines tab.

Create New Routine

Create maintenance routines to facilitate your startup procedure, shutdown procedure,
troubleshooting, or calibration. Ensure that the maintenance routine you create is consistent
with the Luminex analyzer operations and maintenance requirements. See the Luminex
System Manual for more information.

1. Open the Maintenance page, then open the Cmds & Routines tab.

2. In the Routine Name list, click None.

3. In the Commands section, click one or more commands to add to the routine. These

commands display in the plate image and in the Command Sequence list.

4. To change a location (well or reservoir) for a command, click the command in the

Command Sequence list, then click the new location in the plate image.

NOTE: If you attempt to place two or more incompatible commands in the

same well, a message will display instructing you to change the
location of the command. Some commands can be run from the
same location, for example, multiple washes can be run from the
reservoir.

xPONENT

®

3.1 Rev. 2

118

Advertising