Sample volume, Sample dilution, Reagents – Luminex xPONENT 3.1 Rev 2 User Manual

Page 18: Gating, Sample volume sample dilution reagents gating

Advertising
background image

TABLE 2.

Reporter Fluorchromes Wavelengths

R-Phycoerythrin

Alexa 532

Formula weight (Daltons)

240,000

470

Absorbance max (nm)

480, 546, 565

531

Extinction max (M-1cm-1)

1,960,000

83,800

Emission max (nm)

578

554

Quantum yield

0.82

0.8

Sample Volume

Sample volumes or sample sizes range from 10 to 200 µL. Ensure that some sample remains
in the well after aspiration; about 25 µL greater than the sample volume. This amount may
vary depending on the type of plate used. After acquisition, the Luminex analyzer washes the
sample lines resulting in ejection of approximately 165 µL of sheath fluid back into the well for
a 96-well plate. Ensure that there is room to add this amount to the well without overflowing
and contaminating other wells.

The volume restrictions on the assay design can be expounded by the following formula:

Total well volume (µL) – Sample uptake volume (µL) + 165 (µL) <Maximum Well Volume (µL)

• Total well volume = Starting sample volume of a well before the unit samples for

acquisition. Well volume is determined by the consistency of the bead set.

• Sample uptake volume = Uptake volume for acquisition (program this in the protocol as

sample volume).

• 165 (µL) = Volume expelled back as stated in the above paragraph.

• Maximum well volume plate = The maximum volume capacity of the wells in a selected 96-

well microtiter plate.

Sample Dilution

Dilute concentrated biological samples such as plasma or serum, at least 1:5 with reagents
as part of assay setup or as a final dilution step. If running an xMAP-based kit, follow the
dilution instructions in the kit’s product insert.

Reagents

Formulated reagents must be free of particulates other than xMAP beads. Do not dilute
xMAP calibrators or verifiers.

Gating

Gate positions are dependent upon buffer composition. Any changes made to the buffer
composition in an assay may result in a different optimal gate location.

Determine the gating on the Doublet Discriminator channel for the assay during assay
development. The numeric values appear on the left side of the histogram. Use the numerical
gate position, as determined during assay development, to set the gate location in the
protocol.

Introduction

5

Advertising