CT-8000 USER’S MANUAL

REV 2

103

No data or erratic reading on
travel transducer.

• Bad transducer. Perform a transducer diagnostic test using the

instructions in section 4.2.

• Transducer slider is not secured to timing rod.
• Transducer is not secured properly.

No activity on “V2” voltage input
channel.

• The “V2” voltage input channel requires a voltage change greater

than 30V.

How do I turn on automatic
printing of test reports?

• Please see section 3.2.3.