Vanguard CT-8000 User Manual

Page 40

Advertising
background image

CT-8000 USER’S MANUAL

REV 2

35

3.2.8. Configuring the Contact Filter Settings

Although the CT-8000 automatically detects the contact time using its own algorithm, it also
allows the user to enter a custom filter value. The value can be between 1 and 300. A filter
setting of 1 allows the CT-8000 to pick up the first contact transition time after the resistor
contact activity is detected. A filter setting of 300 allows the CT-8000 to pick up the last contact
transition time after the resistor contact activity is detected. Use the steps below to set the
contact filter value:

a. Start from the “START-UP” menu:

Press the

[4]

key (DIAGNOSTICS).

b. The following screen will be displayed:

Press the

[4]

key (NEXT PAGE).

c. The following screen will be displayed:

Press the

[3]

key (CONTACT FILTER).

d. The following screen will be displayed:

Type the new filter value using the numeric keypad and then press the

[ENTER]

key.

The contact filter value will be set and you will be returned to the “START-UP” menu.

NOTE

The filter value will be reset to the automatic setting when the unit’s power is
cycled.

ENTER FILTER SETTING

(1 – 300)

CURRENT SETTING: 1
NEW SETTING?

1. PRINT DATA

2. ENCODER FILTER

3. CONTACT FILTER

1. SLOW CLOSE TEST

2. CHECK HOOKUP

3. TEST TRANSDUCER

4. NEXT PAGE

1. RUN TEST 03/10/10

2. GET RSLT 09:04:05

3. SETUP

4. DIAGNOSTICS