E Series Operator’s Guide

A-6

9650-1210-01 Rev. R

Operating Time

For a new, fully charged battery pack at
20 °C: (68

F)

40 defibrillator discharges at maximum
energy (200 J), or

2.75 hours of continuous ECG
monitoring, or

2.25 hours of continuous ECG
monitoring/pacing at 60 mA, 80 beats per
minute.

For a new, fully charged battery pack at
20 °C: (68

F)

100 defibrillator discharges at maximum
energy (200 J), or

4.25 hours of continuous ECG
monitoring, or

3.75 hours of continuous ECG
monitoring/pacing at 60 mA, 80 beats
per minute.

Standby Life: one month before retest and
recharge.

Note: Each monitoring option added to the E Series device decreases the Operating Time

that can be obtained from a fully charged battery. Refer to the individual option insert
for the operating run time specific to your device. For further details specific to your
device, contact the ZOLL Technical Service Department.

Low Battery
Indicators

Message is displayed on the monitor and a 2-beep, low battery tone sounds once a minute
until just before shutdown, when it beeps twice every 2 seconds. The time from display of
the LOW BATTERY or REPLACE BATTERY prompt until the instrument shuts down varies
depending upon the battery age and condition.