ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 76

Advertising
background image

E Series Operator’s Guide

10-6

9650-1210-01

Rev.

R

If only one parameter is installed on the unit, the trend
screen automatically appears when you press the Trend
softkey.

Not all trending data can be displayed on the monitor at
the same time. However, you can select the resolution of
the time between samples by using the Zoom Softkey.
You can view ALL trending data, or you can view
trending data taken at 5 minute, 10 minute, 15 minute,
30 minute, and 60 minute intervals. Data is presented
with the newest data at the top and the oldest data at the
bottom of the display.

If an alarm occurs while the unit is monitoring vital signs,
such as an NIBP alarm, the unit records the data at the
moment of the alarm, independent of the standard
30 seconds interval recording.

The unit indicates an alarm condition on the trend
display by highlighting the alarmed parameter on the
display. The unit indicates invalid data by substituting a
dashed line (---) for the actual data. Invalid data may
occur, for example, when the measuring probe or device
is not properly connected to the patient.

The most recently recorded trending sample is indicated
by highlighting the time stamp associated with the trend
data. If no time record is highlighted, older data is being
displayed. To view the newest values of recorded data,
press the Newer softkey until the highlighted time
displays.

When viewing the newest data, the display automatically
updates as each new trend record is recorded. To view
older values of recorded data, press the Older softkey
until the desired data displays. If you are viewing the
most recent trending data, the trending screen remains
displayed until you exit the screen. However, if you are
viewing anything other than the most recent trending
data, the trending screen times out three minutes after
the last softkey activation.

Printing a Vital Signs Trend Report

The stripchart prints the trend report using the same
zoom level that the display uses. If the Zoom setting is
set to All and you press the Print softkey on the Trend
display, all trending samples print from the starting
record shown back through to the oldest record.

Data is printed on the stripchart in order of newest data
to oldest data. Alarm conditions are indicated on the
stripchart by placing brackets around alarmed values. As
on the display, invalid data is indicated by a dashed line
(---) for the actual data.

NIBP readings are considered valid for only one trend
record (i.e., the sample during which the NIBP
measurement was completed). All other trend records
indicate invalid NIBP readings.

If the Zoom setting is set to five minutes, when you press
the Print softkey on the Trend display, a reports prints
out that includes only the values sampled at the five
minute intervals.

NIBP Trend Operation

When displaying NIBP trends, only those trending
samples recorded for NIBP readings are displayed.
NIBP History trend reports contain all NIBP records
regardless of the Zoom level.

Print

Retur n

Newer

Zoom

Older

13:25

13:30

120

119

120

100

100

99

4.1

4.3

3.7

36

13

15

TIME

5 MIN

HR/PR SpO2

CO2

EtCO2 RR

09-Jun-10

13:35

13:20

119

99

3.7

15