ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 132

Advertising
background image

E Series Operator’s Guide

B-6

9650-1210-01

Rev.

R

NOT AN E SERIES
DATA CARD

The wrong type of card is
installed. The unit operates, but
no data is being stored.

Insert the correct type of card.

REPLACE CARD

The wrong type of card type is
installed, or the card is defective.
The unit operates, but no data is
being stored.

Erasing the card may allow return
to normal data storage operation

AUDIO NOT
RECORDING

Note: Only possible with the

audio recording option.

Data card has been removed
from the unit while the unit is
turned on. The unit does not
record event or audio
information.

CARD REMOVED

Card was removed from PCMCIA
slot or is not seated properly.

Reinsert card.

EMPTY CARD

Empty data card inserted into
PCMCIA slot, or card was
erased.

Record data on card before
transmission, or insert different
card.

NOT A DATA
CARD

The wrong type of card is
installed.

Insert a different card

VERIFICATION
FAILED

The card was not erased
successfully and is defective.

Insert a different card.

HOST FAULT

UPLOAD
ABORTED

Remote device aborted
transmission. Ensure remote
device is on and ready to accept
data.

Transmission aborted because
the host system detected an error
or user cancelled transmission
via communication program.

Check communication package
on the remote device or PC for
the source of the error. Retry.

Message

Possible Causes and
Corrective Action

SERIAL FAULT or

UPLOAD
ABORTED

E Series unit aborted
transmission. Data retrieval
pre-empted by other E Series
operations (such as alarm
conditions or defibrillation
activity); unit could not retrieve
data from memory.

Wait for E Series events to clear
and retry.

Ensure remote device is on and
ready to accept data.

Check communication settings
on unit and on the host system to
ensure they match.

For serial data transmission:

Check and secure the ZOLL
RS 232 data transfer cable at
the back of the unit, the ZOLL
cable connection to the
RS-232 cable, as well as the
RS-232 cable connection at
the back of the PC.

For Bluetooth data transmission:

Move the host system closer
to the E Series unit and retry.

CARD
PREVIOUSLY
UPLOADED

User inserted a valid data card
that was already uploaded
successfully. Data on card not
erased.

Use a different data card, erase
the data on the card you are
using, or resend the data again.