Synchronized Cardioversion

9650-1210-01 Rev. R

6-3

Once energy is delivered, the display simultaneously
shows the XXXJ DELIVERED and DEFIB XXXJ SEL.
messages. After approximately 5 seconds the XXXJ
DELIVERED
message disappears, and the DEFIB XXXJ
SEL.
message remains to indicate the selected energy
level.

If additional countershocks are necessary, re-adjust the
energy level as necessary, press the SYNC softkey and
repeat. Note the SYNC XXXJ SEL. message must be
displayed prior to pressing the CHARGE button.

If it is necessary to disarm the charged defibrillator (if
countershock is not needed), turn the selector switch to
MONITOR or change the selected energy level. Any
stored energy is discharged internally by the defibrillator.

If the ANALYZE button is pressed while the unit is in
SYNC mode, the unit displays DISABLE SYNC and
disallows ECG rhythm analysis until the unit is taken out
of SYNC mode.

If the defibrillator is not discharged within 60 seconds
after reaching the selected energy level, it automatically
disarms itself. During the ten seconds prior to this
internal disarm, the charge ready tone beeps
intermittently. The charge ready tone then stops and the
defibrillator remains in SYNC mode.

Troubleshooting

If your E Series unit does not function as expected, see
the Defibrillator Troubleshooting section starting on
page 13-5.