ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 32

Advertising
background image

E Series Operator’s Guide

2-8

9650-1210-01

Rev.

R

and pacing current are also recorded. If async pace is active, the annotation “ASYNC PACE” is recorded. AED units
additionally include shock count and AED mode annotations.

Analyze Format

The summary report function records six seconds of pre analysis ECG and 12 seconds of ECG recorded during the
ECG analysis interval with the annotation “SHOCK ADVISED” or “NO SHOCK ADVISED.” AED units additionally
include shock count and AED mode annotations. The date/time printed on the top of the strip corresponds to the ECG
data occurring 6 seconds before the event.

The following annotations may also appear at the top of the Analyze Format printout:

Annotation

Description

POOR PAD CONTACT

The MFE Pads are detected as having poor connection.

ANALYSIS HALTED

The ECG analysis is halted due to either the ANALYZE button being pushed or a fault
condition.

NOISY ECG

Excessive noise is detected.

SHOCK ADVISED

Shockable rhythm has been detected at the end of user-initiated ECG analysis.

NO SHOCK ADVISED

No shockable rhythm has been detected at the end of user-initiated ECG analysis.

ECG TOO LARGE

The amplitude of the ECG signal is too large for proper rhythm analysis.

Actual
event
date/time

Actual
event
date/time