ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 88

Advertising
background image

E Series Operator’s Guide

12-2

9650-1210-01

Rev.

R

3. Turn the defibrillator back on to the selected mode of

operation.

Note: If a battery has been discharged below its safe

limit (8.5 V), which could occur during long-term
storage, the battery fault indicator illuminates
when it is installed. In this case, the battery is not
usable; remove it immediately. Replace with a
fully-charged battery. (In emergency situations,
you can use AC power.)

If the unit is set to PACER mode, pacing may resume
immediately after battery replacement. If this is not
desired, then turn the unit off for more than 10 seconds
prior to replacing the battery.

When operation of the unit is resumed subsequent to
battery replacement, you must re-verify the unit’s
settings (for example, alarms, lead, pacing amplitude
and rate).

Charging and Testing Battery Packs

ZOLL batteries are designed to be charged in the
E Series device or accessory chargers designed for use
with ZOLL devices. ZOLL recommends that you always
have a ZOLL auxiliary battery charger available in order
to charge spare batteries and perform periodic battery
testing; the ZOLL Base Power Charger 4x4 or
SurePower Charger was designed specifically for this
purpose.

With the E Series unit plugged in to AC mains and turned
off, the device recharges the sealed lead acid battery to
greater than 90% capacity within 4 hours, and the
lithium-ion battery pack to greater than 90% capacity in
less than 7 hours. With the E Series unit plugged in and
in use, the device recharges a fully depleted sealed lead
acid battery, or lithium-ion battery pack within 24 hours.

When the E Series products are plugged into AC mains,
the CHARGER ON indicators operate in the following
manner:

The orange-yellow CHARGER ON indicator
illuminates continuously whenever the device is
turned OFF and charging the battery or turned ON
with a battery installed.

The green CHARGER ON indicator illuminates
continuously whenever the unit is turned OFF and
the installed battery has been fully charged to present
capacity.

The green and orange-yellow CHARGER ON
indicators illuminate alternately when no battery is
installed in the unit, or when a battery charging fault
has been detected.

When the device is not connected to AC mains, the
CHARGER ON indicators remain extinguished.

Top Panel Battery LEDs

The top panel of the unit contains 4 LEDs to indicate the
battery charge level and 1 LED to indicate a battery fault.
These LEDs are active only if a SurePower battery is

installed in the unit. If a fault is detected in the battery,
the fault LED is illuminated and the battery charge level
LEDs are turned off. If the fault LED is not illuminated,
the battery charge level LEDs are illuminated as follows:

Troubleshooting

If your E Series unit displays a BATTERY FAULT XX
message, a battery-related fault condition has occurred.
Attempt to clear the fault by removing and reinstalling the
battery pack. Replace the battery and contact the ZOLL
Technical Service department if the condition persists.

Number of
LEDs Lit

Indicates Remaining Runtime of:

4

> 90 minutes

3

> 60 minutes

2

> 30 minutes

1

> 15 minutes

0

< 15 minutes

Advertising