AC Charger
Symptom
Recommended Action
The green and orange-yellow CHARGER ON indicators
illuminate alternately.
•
Verify battery is installed.
•
Turn unit ON to identify the fault condition.
•
Replace battery pack with a fully charged battery pack.
•
If problem persists, replace battery pack, unplug device
from AC mains and plug device back into AC mains.
Unit displays the LOW BATTERY message when it is
plugged into AC mains.
•
Replace battery pack with a fully charged battery pack.
•
Unplug device from AC mains and plug device back into
AC mains.
•
Verify AC mains is working properly.
None of the CHARGER ON indicators are illuminated
when the device is plugged into AC mains.
•
Unplug device from AC mains and plug device back into
AC mains.
•
Verify AC mains is working properly.