Ecg analysis algorithm accuracy – ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 125

Advertising
background image

9650-1210-01 Rev. R

A-27

ECG Analysis Algorithm Accuracy

Sensitivity and specificity are expressions of ECG
analysis algorithm performance when compared to ECG
interpretation by a clinician or expert. Sensitivity refers to
the algorithm’s ability to correctly identify shockable
rhythms (as a percentage of the total number of
shockable rhythms). Specificity refers to the algorithm’s
ability to correctly identify non-shockable rhythms (as a
percentage of the total number of non-shockable
rhythms). The data in the following table summarizes the
accuracy of the ECG analysis algorithm as tested
against ZOLL’s ECG Rhythm Database.

The algorithm sequence takes approximately 9 seconds
and proceeds as follows:

Divides the ECG rhythm into three-second segments.

Filters and measures noise, artifact, and baseline
wander.

Measures baseline content (‘waviness’ at the correct
frequencies frequency domain analysis) of signal.

Measures QRS rate, width, and variability.

Measures amplitude and temporal regularity (‘auto-
correlation’) of peaks and troughs.

Determines if multiple 3 second segments are
shockable then displays SHOCK ADVISED
message.

Clinical Performance Results

The performance of the incorporated analysis algorithm
in a single analysis sequence satisfies the applicable
requirements specified in ANSI/AAMI DF80 (section
6.8.3) and the recommendations by Kerber et al.
(Circulation. 1997;95(6):1677).

Table A-3.

Clinical Performance Results

References:

Young KD, Lewis RJ: “What is confidence? Part 2:
Detailed definition and determination of confidence
intervals”. Annals of Emergency Medicine, September
1997; 30; 311-218

William H. Beyer, Ph.D.: “CRC Standard Mathematical

Tables 28

th

Edition,” CRC Press, Inc, Boca Raton, FL.,

1981, Percentage Points, F-Distribution Table, pg 573.

Rhythms

Sample
Size

Performance Goals Observed

Performance

90% One-sided
Lower Confidence
Limit

Shockable

Sensitivity

Coarse VF

Rapid VT

536

80

>90%

>75%

>95%

>98%

>97%

>96%

Non-shockable

Specificity

NSR

AF, SB, SVT, Heart block,
idioventricular, PVCs

Asystole

2210

819

115

>99%

>95%

>95%

>99%

>99%

>98%

>99%

>99%

>97%

Intermediate

Fine VF

Other VT

69

28

Report only

Report only

Sensitivity

>88%

>96%

>80%

>84%

Advertising