12 lead monitor, Transmitting 12-lead patient records, Troubleshooting – ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 131: Patient information, Press return to exit to 12 lead monitor mode

Advertising
background image

9650-1210-01 Rev. R

B-5

3. To deactivate automatic data transmission, display

the Summary menu (by pressing the SUMMARY
button) and press the Data Relay Off softkey

Note: If you change the destination port while a record

is already being transmitted, that transmission
continues to the previously selected port. All other
records in the queue will be transmitted to the
newly-selected port.

Note: If you change ports from Bluetooth to RS-232,

and then you want to change back to Bluetooth,
the host system must reestablish a connection
with the E Series unit. However, power to the
Bluetooth module is not turned on again until a
record starts transmitting out the Bluetooth port.

Transmitting 12-Lead Patient Records

You can transmit 12-lead ECG patient records to a
remote host system running ZOLL Data Relay software
without placing the E Series unit into Realtime mode.

1. With the E Series unit in Monitor mode (ON for AED

units), press the 12 Lead softkey to access the 12
Lead menu.

2. Press the PT Info softkey to access the Patient

Information menu.

3. Press the Patient Records softkey to display the

Patient Records menu.

4. Select a Patient record and press the Transmit

softkey.

5. Select the Data Relay option and press the Dial

Phone# softkey.

If you are transmitting via Bluetooth, then the green
power LED on the E Series unit illuminates. You must
now establish a connection between the host system
and the E Series unit.

Troubleshooting

If configured, the unit displays the following messages
when it encounters problems during data transmission.

01:20

12 LEAD MONITOR

ECG

24

NIBP

100

120

80

SpO2%

98

CO2

38

m

m

Hg

24

RR

ECG

II

120

x2

Acquire

Settings

Alar ms

Retur n

PT Info

01:22

24

NIBP

100

120

80

SpO2%

98

CO2

38

m

m

Hg

24

RR

ECG

II

120

x2

ID #

Age

Patient

Records

Retur n

Gender

PATIENT INFORMATION

:

Press Return to Exit to

12 Lead Monitor Mode

60

Male

Message

Possible Causes and
Corrective Action

CARD FULL

Data card is full. No more data
can be stored on the card, but the
unit continues to operate. This
prompt is only issued when the
unit is not analyzing or charging.

Insert another card, or print event
data on stripchart recorder.

INSERT CARD

Card is not installed or not seated
properly in the unit.

CARD NOT
ERASED
PROPERLY /
ERASE CARD

An Erase Card operation was
interrupted, or the card is
damaged. The unit operates, but
no data is being stored.

Erasing the card may allow return
to normal data storage operation.

CARD WRITE-
PROTECTED

The card is write-protected. The
unit operates, but no data is
being stored.

Adjust the write-protect tab on
the card to allow writing.

14:07:32

Print

Record

Print

All

17 JUN 05

17 JUN 05
17 JUN 05

200506171407

14:07:32

11:52:48

200506171152

17 JUN 04

200506171324

13:24:03