ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 112

Advertising
background image

E Series Operator’s Guide

A-14

9650-1210-01 Rev. R

Table A-1.

Delivered Energy at Every Defibrillator Settings into a Range of Loads

Figures A-1 through A-20 show the rectilinear biphasic waveforms that are produced when the E Series defibrillator is
discharged into loads of 25, 50, 75, 100, 125, 150, and 175 ohms at each energy setting (200, 150, 120, 100, 85, 70,
50, 30, 20, 15, 10, 9, 8, 7, 6, 5, 4, 3, 2 and 1 joule[s]).

The vertical axis shows the current in amperes (A); the horizontal axis shows the duration in milliseconds (ms).

Selected

Energy

Load

25

50

75

100

125

150

175

Accuracy

1 J

1 J

1 J

1 J

1 J

1 J

1 J

1 J

±3J

2 J

1 J

2 J

3 J

2 J

2 J

2 J

2 J

±3J

3 J

2 J

3 J

4 J

4 J

3 J

3 J

3 J

±3J

4 J

3 J

4 J

5 J

5 J

5 J

4 J

4 J

±3J

5 J

4 J

5 J

6 J

6 J

6 J

5 J

5 J

±3J

6 J

4 J

6 J

8 J

7 J

7 J

6 J

6 J

±3J

7 J

5 J

7 J

9 J

9 J

8 J

8 J

7 J

±3J

8 J

6 J

8 J

10 J

10 J

9 J

9 J

8 J

±3J

9 J

7J

9 J

11 J

11 J

10 J

10 J

9 J

±3J

10 J

7 J

10 J

13 J

12 J

12 J

11 J

10 J

±3J

15 J

11 J

15 J

19 J

19 J

17 J

16 J

15 J

±3J

20 J

15 J

19 J

25 J

25 J

23 J

22 J

20 J

±3J

30 J

22 J

29 J

38 J

37 J

35 J

32 J

30 J

±15%

50 J

37 J

48 J

63 J

62 J

58 J

54 J

50 J

±15%

70 J

53 J

69 J

90 J

89 J

83 J

77 J

71 J

±15%

85 J

65 J

86 J

112 J

110 J

103 J

96 J

89 J

±15%

100 J

74 J

97 J

126 J

125 J

116 J

108 J

100 J

±15%

120 J

88 J

116 J

151 J

149 J

139 J

129 J

120 J

±15%

150 J

110 J

145 J

188 J

186 J

174 J

161 J

150 J

±15%

200 J

146 J

214 J

238 J

234 J

213 J

195 J

179 J

±15%