Warning – ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 37

Advertising
background image

9650-1210-01 Rev. R

3-1

SECTION 3
MANUAL DEFIBRILLATION

Emergency Defibrillation Procedure with Paddles or MFE Pads

You can perform manual defibrillation using either
paddles or MFE pads; the manual defibrillation
procedure described in this section allows for either.

Before you begin manual defibrillation:

Determine the patient’s condition following medical
protocols and by verifying:

Unconsciousness

Absence of breathing, and

Absence of pulse

If appropriate, begin CPR following medical protocol,
and request additional assistance.

Note: If you are using MFE pads, prepare the patient

and attach the MFE pads as described in “MFE
Pad Application/Connection” on page 1-9.

If connecting to PADS through the AutoPulse Plus, refer

to the User Guide AutoPulse

®

Resuscitation System

Model 100 with Defibrillator Interface addendum (part
number 9650-0720-01) for instructions on properly
connecting the E Series to the AutoPulse Plus. It is
recommended that the user cycle through the lead
setting until the APLS icon appears in the upper right-
hand corner of the display, indicating that the unit
recognizes the connection to the AutoPulse Plus.

1

Prepare the unit for defibrillation.

Turn the unit to DEFIB.

The unit automatically defaults to 120 J or the first shock
energy selection configured by the user. For information
about configuring energy levels, refer to the E Series
Configuration Guide
(part number 9650-1201-01).

Paddles/pads are a defibrillation-protected Type BF patient connection.

ECG leads are a defibrillation-protected Type CF patient connection.

WARNING

To avoid risk of electrical shock, do not allow electrolyte gel to accumulate on hands
or paddle handles.

When defibrillating with paddles, use your thumbs to operate the SHOCK buttons in
order to avoid operator shock. No portion of the hand should be near the paddle
plates.

The AutoPulse Plus is only intended for use on adults 18 years of age or older.
When using the E Series and AutoPulse Plus as a system, this age restriction
applies to the E Series as well.

WARNING

If the APLS icon does not appear, check that the
E Series and AutoPulse Plus are properly connected. If
the APLS icon still does not appear, or a
PADDLE FAULT occurs, discharge energy internally by
changing the energy selection, disconnect the
Multi-Function Cable and PADS from the AutoPulse
Plus, and connect the Multi-Function Cable directly to
the PADS.

0!#%2

/&&