Recorder – ZOLL E Series Monitor Defibrillator Rev R User Manual
Page 93

Troubleshooting Guidelines
9650-1210-01 Rev. R
13-3
Recorder
Symptom
Recommended Action
Unit displays the CHECK RECORDER prompt.
•
Ensure the recorder door is shut properly.
•
Ensure adequate paper supply in recorder paper tray
(see “Changing Recorder Paper” on page 11-4).
•
Remove paper, check paper type, check recorder for
paper jam, and then reload paper.
Recorder makes stuttering sound when activated.
•
Check recorder for paper jam.
Light or poor quality printing is observed.
•
Ensure correct paper type (ZOLL P/N 8000-0300) is in
use.
•
Ensure paper is installed grid-side against recorder print
head.
•
Ensure that door is latched shut properly. You must
listen for clicks on both corners of the door when
latching the door shut.
•
Clean recorder print head (see “Cleaning the Recorder
Printhead” on page 11-1).
Summary report does not print when you press
SUMMARY button.
•
Wait 15 seconds and try again. (The unit requires 15
seconds after an event occurs to finish storing a record
of the event. A summary report cannot be printed during
this time.)