Troubleshooting Guidelines

9650-1210-01 Rev. R

13-3

Recorder

Symptom

Recommended Action

Unit displays the CHECK RECORDER prompt.

Ensure the recorder door is shut properly.

Ensure adequate paper supply in recorder paper tray
(see “Changing Recorder Paper” on page 11-4).

Remove paper, check paper type, check recorder for
paper jam, and then reload paper.

Recorder makes stuttering sound when activated.

Check recorder for paper jam.

Light or poor quality printing is observed.

Ensure correct paper type (ZOLL P/N 8000-0300) is in
use.

Ensure paper is installed grid-side against recorder print
head.

Ensure that door is latched shut properly. You must
listen for clicks on both corners of the door when
latching the door shut.

Clean recorder print head (see “Cleaning the Recorder
Printhead” on page 11-1).

Summary report does not print when you press
SUMMARY button.

Wait 15 seconds and try again. (The unit requires 15
seconds after an event occurs to finish storing a record
of the event. A summary report cannot be printed during
this time.)