Electromagnetic Immunity Declaration (EID)

Immunity Test

IEC 60601 Test Level

Compliance Level

Electromagnetic Environ-
ment — Guidance

Electrostatic discharge
(ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, con-
crete or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.

Electrical fast
transient/burst
IEC 61000-4-4

± 2 kV for power supply
lines

± 1 kV for input/output
lines

± 2 kV for power supply
lines

Not applicable

Mains power quality should be
that of a typical commercial or
hospital environment.

Surge
IEC 61000-4-5

± 1 kV differential mode

± 2 kV for common
mode

± 1 kV differential mode

± 2 kV for common
mode

Mains power quality should be
that of a typical commercial or
hospital environment.

Voltage dips, short
interruptions and volt-
age variations on power
supply input lines
IEC 61000-4-11

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in U

T

)

for 5 sec.

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in U

T

)

for 5 sec.

Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the E Series unit requires
continued operation during
power mains interruptions, it is
recommended that the E Series
unit be powered by an uninter-
ruptible power supply or a bat-
tery.

Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic
fields should be at levels char-
acteristic of a typical location in
a typical commercial or hospital
environment.

Note: U

T

is the AC mains voltage prior to application of the test level.