Printing a report, Printing a call report, Printing a partial summary report – ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 33: Printing an incident log, Operating controls and indicators, Manual mode activated

Advertising
background image

Operating Controls and Indicators

9650-1210-01 Rev. R

2-9

Manual Mode Activated

AED versions of the E Series will record a “MANUAL
MODE STARTED” summary report event when the
device is switched from Semiautomatic mode to Manual
mode.

Printing a Report

To print the stored information, press the SUMMARY
button below the screen display. Then press the
corresponding softkey to print a “call” report, summary
report (print chart), or incident log.

The recorder prints all summary report events currently
in memory in chronological order. If the unit is equipped
with the 12-lead option, the recorder will print all 12-lead
patient records currently in memory at the end of the
summary report printout. If the stripchart recorder is on
or the defibrillator is charged, summary report printing is
disabled. In addition:

To stop printing a report, press the RECORDER
button or turn the unit off. You can print an unlimited
number of copies of the report by pressing the
SUMMARY button and corresponding print softkey
again.

If you press the RECORDER button while printing a
summary report, the unit stops printing the report.
Press the RECORDER button again to begin printing
an ECG trace. The stripchart recorder runs
continuously until the button is pressed again.

If you press the SUMMARY button and a
corresponding print softkey while a report is already
printing, the current report stops printing and a new
report begins to print.

Printing is interrupted if a vital sign alarm occurs (i.e.
HR, SpO2, etc.), the ANALYZE button is pressed, or
the defibrillator is charged.

If the recorder is out of paper when the SUMMARY
button and a corresponding print softkey are pressed,
a CHECK RECORDER message appears on the
display. Load paper and press the SUMMARY button
again to select the report to print.

Printing a Call Report

A Call Report is an abbreviated summary report
containing only those events associated with a specific
call or run. To print a call report:

1. Press the SUMMARY button.
2. Press the Print Call softkey.

3. Use the scroll keys on top of the unit to scroll through

the list of start times associated with different calls.

4. Press the Commit ( ) key on top of the unit, or the

Print Records softkey, to print the events associated
with the selected call.

Printing a Partial Summary Report

To print only a portion of the Summary Report:

1. Press the SUMMARY button.
2. Press the Print Chart softkey.
3. Press the Print Range softkey.
4. Use the scroll keys on top of the unit to scroll through

the list of events.

5. Press the Commit ( ) key on top of the unit, or the

Print Records softkey, to print the specified event
and all subsequent events.

Printing an Incident Log

An incident log is an abbreviated list of all major events
recorded in the summary report. You can print out an
incident log that includes the time of occurrence of the
following events:

E Series unit powered on

Defibrillation advisory messages (for example,
CHECK PATIENT and SHOCK ADVISED)

Defibrillation shocks (including energy level)

Pacer mode activated

#(!2'%2

35--!29

#/$%

-!2+%2

Print

Call

Print

Chart

Print

Log

.)"0

Realtime

Start

Trend

24

NIBP

-

---

---

SpO2%

---

CO2

---

m

m

Hg

---

RR

ECG

PADS

70

x1



kPa



Abor t

Print

Records

02NOV 11:43:11 Call Started
02NOV 11:40:05 Call Started



02NOV

3952 Call Started

02NOV 11:38:11 Call Started
02NOV 11:35:05 Call Started

24

NIBP

-

---

---

SpO2%

---

CO2

---

m

m

Hg

---

RR

ECG

PADS

70

x1



kPa



Abor t

Print

Records

02NOV 14:18:02 Intubate
02NOV 14:12:29 CPR



02NOV 11:52:48 RECORDED ECG
02NOV 11:06:14 POWER ON

02NOV 11:54:03 DEFIB ADVISORY

Advertising