Real cpr help field, Warning – ZOLL E Series Monitor Defibrillator Rev R User Manual
Page 57
9650-1210-01 Rev. R
7-1
SECTION 7
REAL CPR HELP
When used with CPR-D-padz, the E Series unit can 
provide rescuers with feedback about the quality of CPR 
they are delivering to their patients. The way in which 
feedback is provided varies with respect to the 
operational mode and user configuration, but is derived 
from compression rate and depth measurement. The 
CPR Dashboard™ contains additional features such as 
the Compression Release Bar.
When applied according to package instructions, ZOLL 
CPR-D-padz provide a chest compression sensor that is 
located between the rescuer's hands and the patient's 
lower sternum. This sensor monitors the rate and depth 
of chest compressions and sends this information to the 
E Series unit for processing and display. 
The E Series defibrillator uses this information to provide 
feedback to the rescuer in one or more of the following 
forms:
•
CPR Index™
•
CPR Idle Time Display
•
CPR Rate Metronome
•
Voice prompts
•
Chest Compressions Bar Graph
•
Informational Messages
Prepare the Patient and Attach the 
CPRD-to-MFC Connector
Prepare the patient as described on page 1-9. Attach the 
CPR-D-padz to the patient according to the instructions 
on the electrode packaging. Connect the Multifunction 
cable to the narrow end of the CPRD-to-MFC connector 
(if not already connected). Connect the CPR-D-padz to 
the wide end of the CPRD-to-MFC connector.
Ensure that the CPR-D-padz are making good contact 
with the patient’s skin. If the pads are not making good 
contact, the messages CHECK PADS and POOR PAD 
CONTACT will be displayed, and energy will not be 
delivered. If a short circuit exists between the pads, the 
message DEFIB PAD SHORT will be displayed.
Note: If the E Series displays the message CABLE
FAULT or PADDLE FAULT, energy will not be 
delivered. Check the connection to the 
CPRD-to-MFC connector. If the message does 
not clear, disconnect the CPRD-to-MFC 
connector and connect the cable directly to the 
CPR-D-padz. If the fault clears, CPR feedback 
will not be available, but the device will be able to 
deliver energy.
Real CPR Help Field
Whenever CPR-D-padz are connected to the E Series 
defibrillator and the CPR Dashboard is configured off, 
the unit illuminates the Real CPR Help field in the upper 
right side of the display. This field includes the CPR 
The CPR sensor is defibrillation-proof Type BF equipment.
WARNING
•
The Real CPR Help
®
function is not intended for use on patients under 8 years of age.
•
The CPRD-to-MFC connector is intended for use with the E Series, or other ZOLL defibrillators where indicated.
•
Real CPR Help is disabled when using CPR-D-padz through an AutoPulse Plus.