ZOLL E Series Monitor Defibrillator Rev R User Manual
Page 67
9650-1210-01 Rev. R
9-1
SECTION 9
NONINVASIVE TEMPORARY PACING 
(PACER VERSION ONLY)
Noninvasive Temporary Pacing
Some E Series products contain a VVI demand 
pacemaker — a safe and effective design for 
Noninvasive Temporary Pacemakers. Proper demand 
pacing requires a reliable, high-quality surface ECG 
signal. 
Determine patient condition and provide care 
following medical protocols.
Prepare Patient
Remove all clothing covering the patient’s chest. Dry 
chest if necessary. If the patient has excessive chest 
hair, clip it to ensure proper adhesion of electrodes. 
1
Apply Electrodes and/or MFE Pads
Connect the electrodes to the ECG cable leads, and the 
ECG cable to the E Series unit. Apply ECG electrodes 
(refer to “ECG Monitoring” on page 10-1). Adjust the 
ECG size and lead for a convenient waveform display. 
Verify proper R-wave detection. The heart-shaped 
symbol flashes with each R-wave when proper detection 
is taking place.
Attach and connect the MFE pads
This is described in “MFE Pad Application/Connection” 
on page 1-9.
2
Turn Selector Switch to PACER
Set Pacer Output to 0 mA
If the unit has just been turned on, the PACER OUTPUT 
is automatically set to 0 mA.
3
Set Pacer Rate
Set PACER RATE to a value 10-20 ppm higher than 
patient’s intrinsic rate. If no intrinsic rate exists, use 
100 ppm.
Paddles are a defibrillation-protected Type BF patient connection.
ECG leads are a defibrillation-protected Type CF patient connection.
WARNING
•
To avoid risk of electrical shock, do not touch the gelled area of the MFE Pads while pacing.
•
MFE Pads should be replaced after 8 hours of continuous pacing (2 hours for Radiolucent stat-padz) to ensure 
maximum patient benefit. 
•
Pacing with CPR-D-padz is not recommended. For the most effective pacing, use MFE pads.
•
Prolonged pacing (in excess of 30 minutes), particularly in neonates or adults with severely restricted blood flow, 
may cause burns. Periodic inspection of the underlying skin is recommended.
•
If the unit was not turned off and less than 10 minutes have elapsed since the pacing mode was last used, 
reactivating the pacer mode causes pacing to immediately resume at the previously selected mA and ppm 
settings.
•
Pacing must be turned off before defibrillating with a second defibrillator. Otherwise, the E Series unit may be 
damaged.
0!#%2
/&&
-/.)4/2
$%&)"