General maintenance – ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 81

Advertising
background image

General Maintenance

9650-1210-01 Rev. R

11-3

5. When charge ready tone sounds, use the defibrillator

ENERGY SELECT buttons on either the sternum
paddle or defibrillator front panel to change the
selected energy to 20 joules.

6. Verify that the unit disarms itself.
7. Use the defibrillator ENERGY SELECT buttons on

either the sternum paddle or defibrillator front panel to
change the selected energy back to 30 joules.

MFC Setup:

1. Plug the MFC cable into the unit, making sure it is not

plugged into the test connector.

2. Set the selector switch on the front panel to DEFIB

and select 30 J.

3. Verify that the unit alternately displays the CHECK

PADS and POOR PAD CONTACT messages.

4. Insert the end of the Multi-Function Cable into its test

connector (Attached to the Multi-function cable), as
shown below.

5. Verify that the unit displays the DEFIB PAD SHORT

message.

Energy Delivery Test

1. Press the CHARGE button on the front panel.
2. Wait for the charge ready tone to sound and verify

that the energy ready value displayed on the monitor
registers 30 joules (DEFIB 30J READY).

3. Press and hold the SHOCK button on the front panel

of the defibrillator (or the shock buttons on the
paddles) until discharge occurs.

4. Verify that the units displays the TEST OK message

and prints a stripchart of the event, noting the energy
delivered and impedance data.

Note: If TEST FAILED appears, contact the ZOLL

Technical Service Department immediately.

Testing the Pacer (Pacer Version Only)

Perform these tests on all pacer-equipped units
periodically.

1. Connect the ECG leads and Multi-function cable to

the simulator.

2. Turn the selector switch to PACER.
3. Turn the pacer rate control to 150 ppm.

4. Press the RECORDER button to generate a strip,

and then again to stop printing.

5. Verify that the pacing stimulus markers (

) occur

approximately every 10 small divisions (2 large
divisions or 1 cm).

6. Press the RECORDER button to generate another

stripchart, then press and hold the 4:1 button on the
front panel.

7. Press the RECORDER button again to stop printing.
8. Verify that the frequency of the markers decreases (8

large divisions or 4 cm between each marker).

9. Turn the PACER OUTPUT control to 0 mA.
10. Verify there are no CHECK PADS or POOR PAD

CONTACT messages.

11. Disconnect MFE Pads or paddles from the

Multi-Function cable, and slowly turn the knob up to
change the pacer output to 16 mA.

12. Verify that the Pace alarm sounds and a CLEAR

PACE ALARM prompt flashes.

13. Connect the Multi-Function cable to the test

connector, and press the Clear Pace Alarm softkey.

14. Verify that the unit Pace alarm stops and that the unit

removes error messages from the LCD.