Yokogawa Button Operated MV2000 User Manual

Page 112

Advertising
background image

4-12

IM MV1000-01E

Explanaton

• Operatons That Begn at the Same Tme as Memory Samplng (Memory

start)

• Waveform display updating on the trend display
• Reporting (/M1 and /PM1 options)
• Computation functions (/M1 and /PM1 options), if they are configured to do so.

For details, see section 10.4.

• Operatons That Stop at the Same Tme as Memory Samplng (Memory

stop)

• Waveform display updating on the trend display
• Reporting (/M1 and /PM1 options)
• Computation operations (/M1 and /PM1 options), if they are being performed.

• Save Problems

The following problems may arise when the MV is continually accessing internal or

external memory. If these problems arise, the storage media access indicator will light
frequently.
• External storage media file save error
• Access timeout
These problems can be dealt with in the following ways.
• If you are using the event action function and saving small consecutive files,

increase the size of the files being saved.

• If you are saving multiple files to the same directory on an external storage

medium, change the name of the Save directory, making sure that the number of
files in the directory does not exceed 1000.

• If data recording and display are using up resources (for example if you are

recording on multiple channels using a fast sampling rate, and displaying four trend
displays on the 4-panel display), use a slower sampling rate or change the display.

4.4 Startng/Stoppng Recordng and Savng Measured Data

Advertising