6 general specifications, 6 general specifications -14, 6 general specfcatons – Yokogawa Button Operated MV2000 User Manual

Page 282: Constructon, Normal operatng condtons, Input termnal (nput termnal suffx code -1), Input termnal (nput termnal suffx code -2)

Advertising
background image

13-14

IM MV1000-01E

13.6 General Specfcatons

Constructon

Item

Specfcatons

Material

Case: Aluminum plate (MV2000), metal plate (MV1000)
Bezel: Polycarbonate

Color

Case: Smoke blue (Munsell 4.1PB6.0/4.5 or equivalent)
Bezel: Light gray (Munsell 5.2PB8.2/1.0 or equivalent)

External dimensions

MV1000: 189(W) × 177(H) × 253(D) mm, 189(W) × 177(H) × 259(D) mm

*1

, 189(W) × 186(H) ×

259(D) mm

*2

MV2000: 307(W) × 273(H) × 254(D) mm , 307(W) × 273(H) × 260(D) mm

*1

*1 With clamp terminals (input terminal suffix code is -1) or pulse inputs (/PM1 option)
*2 Rechargeable battery model (power supply suffix code is -3)

Weight

MV1004, MV1006, MV1012: Approx. 3.3 kg, Approx. 4.5 kg

*1

, MV1008, MV1024: Approx. 3.5 kg,

Approx. 4.7 kg

*1

MV2008, MV2020: Approx. 5.3 kg, MV2010: Approx. 5.2 kg, MV2030: Approx. 5.5

kg, MV2040, MV2048: Approx. 5.6 kg excluding options
*1 Rechargeable battery model (power supply suffix code is -3)

Normal Operatng Condtons

Item

Specfcatons

Supply voltage

90 to 132, 180 to 250 VAC

Power supply frequency

50 Hz ± 2%, 60 Hz ± 2%

Ambient temperature

0 to 40°C, 10 to 40°C (Rechargeable battery model (power supply suffix code is -3))

Ambient humidity

20 to 80%RH (at 5 to 40°C)

Vibration

10 to 60 Hz, 0.2 m/s

2

Shock

Not allowed

Magnetic field

400 A/m or less (DC and 50/60 Hz)

External noise

Normal mode (50/60 Hz)

DC voltage

The peak value including the signal must be less than 1.2 times the measuring range.

Thermocouple

The peak value including the signal must be less than 1.2 times the measuring thermal

electromotive force.

RTD

50 mV or less

Common mode noise

250 VACrms or less for all ranges (50/60 Hz)

Maximum noise voltage between channels

250 VACrms (50/60 Hz) or less

Installation position

Horizontal
You can tilt the MV by bringing the front legs out.

Warm-up time

At least 30 minutes after power-on

Installation location

Indoors

Operating altitude

2000 m or less

Input Termnal (nput termnal suffx code -1)

Item

Specfcatons

Type

Clamp terminal
Attachable and detachable for each channel
Recommended wire size: 0.08 mm

2

to 1.5 mm

2

(AWG28 to 16)

Input Termnal (nput termnal suffx code -2)

Item

Specfcatons

Type

M4 screw

Advertising
This manual is related to the following products: