10-7

IM MV1000-01E

Computat

on and Report Funct

ons (/M1 and /PM1 opt

ons)

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

Basc Arthmetc

Expresson Example

• Addition 001+002
(Determines the sum of the measured values of channel 1 and channel 2)
• Subtraction 001–002
(Determines the difference between the measured values of channel 1

and channel 2)

• Multiplication 001*K03
(Multiplies the measured value of channel 1 by constant K03)
• Division 001/K02
(Divides the measured value of channel 1 by constant K02)

Power and Other Operatons

Expresson Example

• Power 001**002
(Determines the measured value of channel 1 to the power of the

measured value of channel 2)

• Square root SQR(002)
(Determines the square root of the measured value of channel 2)
• Absolute value ABS(002)
(Determines the absolute value of the measured value of channel 2)
• Common logarithm LOG(001)
(Determines the common logarithm (log

10

) of the measured value

of channel 1)

• Natural logarithm LN(001)
(Determines the natural logarithm of the measured value of

channel 1)

• Exponent EXP(001)
(Determines e to the power of the measured value of channel 1)

Relatonal Computaton

Expresson Example

002.LT.003

If the measured value of channel 2 is less than the measured value of channel 3, the
computed result is 1. Otherwise, the result is zero.

002.GT.003

If the measured value of channel 2 is greater than the measured value of channel 3, the
computed result is 1. Otherwise, the result is zero.

002.EQ.003

If the measured value of channel 2 is equal to the measured value of channel 3, the
computed result is 1. Otherwise, the result is zero.

002.NE.003

If the measured value of channel 2 is not equal to the measured value of channel 3, the
computed result is 1. Otherwise, the result is zero.

002.GE.003

If the measured value of channel 2 is greater than or equal to the measured value of
channel 3, the computed result is 1. Otherwise, the result is zero.

002.LE.003

If the measured value of channel 2 is less than or equal to the measured value of channel 3,
the computed result is 1. Otherwise, the result is zero.

10.2 Wrtng Expressons

Advertising