24 VDC Transmtter Power Supply (/TPS2 and /TPS4)

Item

Specfcatons

Number of loops

2 (/TPS2), 4 (/TPS4)

Output voltage

22.8 to 25.2 VDC (under rated load current)

Rated output current

4 to 20 mADC

Maximum output current

25 mADC (overcurrent protection operation current: approx. 68 mADC)

Allowable conductor resistance

RL ≤ (17.8 – minimum transmitter operation voltage)/0.02 A
where 17.8 V is the result obtained by subtracting the maximum drop voltage of 5 V,

when the load shunt resistance is 250 Ω, from the minimum output voltage of 22.8 V

Max. length of wiring

2 km (when using a CEV cable)

Insulation resistance

20 MΩ or more at 500 VDC between output terminal and ground

Dielectric strength

500 VAC (50/60 Hz, I = 10mA) for one minute between output terminal and ground
500 VAC (50/60 Hz, I = 10mA) for one minute between output terminals

Pulse Input (/PM1)

Item

Specfcatons

Pulse input

Number of Inputs:

3 (8 when using the remote control input terminals)

Input type

Isolated from the main circuitry through a photocoupler and built-in isolated power supply

for the input terminals.
Shared common for pulse inputs

Input type and signal level

Voltage-free

contact

Contact closed at 200 Ω or less and contact open at 100 kΩ or

greater.

Open collector

ON voltage: 0.5 V or less (sink current 30 mA or more), leakage

current when OFF: 0.25 mA or less

Counting

Counts the rising edges of pulses.
For voltage-free contact input: Contact open to contact close
For open collector: The H terminal voltage level changing from high to low

Allowable input voltage

30 VDC

Max. sampling pulse period

100 Hz

Minimum detected pulse width 5 ms or more for both low (closed) and high (open)
Pulse detection period

Approx. 3.9 ms (256 Hz)

Pulse measuring accuracy

±1 pulse

Pulse count interval

Scan interval or 1 s

Miscellaneous

Pulse input terminals can be used as remote control input terminals, isolated from

remote control input terminals

Remote control

Number of inputs: 5. Other items are the same as with remote control (/R1)
Remote control input terminals can be used as pulse input terminals.

Computation Function

Same as the computation function (/M1)

Calbraton Correcton (/CC1)

Item

Specfcatons

Calibration correction method

Corrects the measured value of each channel using segment linearizer approximation
Number of segment points: 2 to 16 (including the start and end points)

External Input Channel Functon (/MC1)

Item

Specfcatons

Function

Loads data from other instruments by using the Modbus client or Modbus master

function and displays, records, and saves the data

Number of channels

240 channels (201 to 440)

Display

Same as the measurement channels

Data saving

Same as the measurement channels

Manual sampling

Specify up to 120 channels from measurement, computation, and external input

channels.