Chapter 13 specifications, 1 signal input and alarm, Chapter 13 specfcatons – Yokogawa Button Operated MV2000 User Manual

Page 269: 1 signal input and alarm -1, 1 sgnal input and alarm, App index, Sgnal input

Advertising
background image

13-1

IM MV1000-01E

Spec

fcat
ons

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

13.1 Sgnal Input and Alarm

Sgnal Input

Item

Specfcatons

Number of inputs, scan interval, and A/D integration time

25 ms

125 ms

4
8
6
12
24
8
10
20
30
40
48

Fast sampling mode*

Normal mode

Number of Inputs:

Model

Scan Interval

A/D Converter Integration Time

100 ms

60 Hz/50 Hz

600 Hz (fixed)

125 ms, 250 ms

1 s, 2 s, 5 s

25 ms

125 ms

125 ms, 250 ms

1 s, 2 s, 5 s

2 s, 5 s

2 s, 5 s

MV1004
MV1008
MV1006
MV1012
MV1024
MV2008
MV2010
MV2020
MV2030
MV2040
MV2048

* Not available on models equipped with external input channels (/MC1 option).

Input types

DC voltage, 1-5V, thermocouple (TC), resistance temperature detector (RTD), ON/OFF input

(DI), and DC current (by adding an external shunt resistor)

Input format

Floating unbalanced input

Measurement range and measurable range

DC voltage

20 mV

–20.000 to 20.000 mV

60 mV

–60.00 to 60.00 mV

200 mV

–200.00 to 200.00 mV

2 V

–2.0000 to 2.0000 V

6 V

–6.000 to 6.000 V

20 V

–20.000 to 20.000 V

50 V

–50.00 to 50.00 V

1-5V

1-5 V

6

0.800 to 5.200 V

Thermocouple R

1

0.0 to 1760.0°C

S

1

0.0 to 1760.0°C

B

1

0.0 to 1820.0°C

K

1

–200.0 to 1370.0°C

E

1

–200.0 to 800.0°C

J

1

–200.0 to 1100.0°C

T

1

–200.0 to 400.0°C

N

1

0.0 to 1300.0°C

W

2

0.0 to 2315.0°C

L

3

–200.0 to 900.0°C

U

3

–200.0 to 400.0°C

WRe

4

0.0 to 2400.0°C

RTD

Pt (Pt100)

5

–200.0 to 600.0°C

JPt (JPt100)

5

–200.0 to 550.0°C

DI

Voltage

0: Less than 2.4 V. 1: 2.4 V or higher (judged at the 6 V range)

Contact

7

0: Open. 1: Closed (parallel capacitance of 0.01 µF or less)

1: R, S, B, K, E, J, T, N: IEC584-1 (1995), DIN IEC584, JIS C1602-1995

2: W: W-5%Re/W-26%Re (Hoskins Mfg. Co.), ASTM E988

3: L: Fe-CuNi, DIN43710, U: Cu-CuNi, DIN43710

4: WRe: W-3%Re/W-25%Re (Hoskins Mfg. Co.)

5: Pt100: JIS C1604-1997, IEC751-1995, DIN IEC751-1996

JPt100: JIS C1604-1989, JIS C1606-1989

Measuring current: i = 1mA (Pt100, JPt100)

6: The range for linear scaling of 1-5V inputs. Burnout detection and low-cut functions are

available.

7:

The detected current value is approx. 10 μA.

32 to 3200°F

32 to 3200°F

32 to 3308°F

–328 to 2498°F

–328.0 to 1472.0°F

–328.0 to 2012.0°F

–328.0 to 752.0°F

32 to 2372°F

32 to 4199°F

–328.0 to 1652.0°F

–328.0 to 752.0°F

32 to 4352°F

–328.0 to 1112.0°F

–328.0 to 1022.0°F

Measurable Range

Range

Input Type

Chapter 13 Specfcatons