Chapter 12 troubleshooting and maintenance, 1 list of messages, Chapter 12 troubleshootng and mantenance – Yokogawa Button Operated MV2000 User Manual

Page 246: 1 list of messages -1, 1 lst of messages, App index

Advertising
background image

12-1

IM MV1000-01E

Troubleshoot

ng and Ma

ntenance

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

12.1 Lst of Messages

There are cases in which error codes and messages are displayed on the screen during
operation. A list of the possible error codes and messages are given in the table below.
Communication error codes and messages are also listed.
Error responses to communication commands are output in English.

Errors Related to Parameter Settngs

• Settng Errors

Code

Message

Explanaton/Countermeasures/Ref. secton

1

System error.

Contact your nearest YOKOGAWA dealer.

2

Incorrect date or time setting.

Enter a correct value.

3

A disabled channel is selected.

Specify a channel that is not set to Skip or Off.
Specify a channel that is installed.

4

Incorrect function parameter.

See the Communication Interface User’s Manual.

5

The input numerical value exceeds the set range.

Enter a proper value.

6

Incorrect input character string.

Enter a proper character string.

7

Too many characters.

Enter the correct number of characters.

8

Incorrect input mode.

Specify a correct mode.
See section 3.3.

9

Incorrect input range code.

Specify a correct range code.
See section 3.3.

11

Range settings are not same within the selected channels.

Specify channels with the same range setting.
See section 3.9.

21

Cannot set an alarm for a skipped channel.

Cannot be specified on channels set to Skip.
See section 3.7.

22

The upper and lower span limits are equal.

Cannot be set to the same value.
See section 3.3.

23

The upper and lower scale limits are equal.

Cannot be set to the same value.
See section 3.3.

24

The lower limit of the span band is greater than the upper

limit.

Set the lower limit less than the upper limit.
See section 3.3.

25

The lower limit of the scale band is greater than the upper

limit.

Set the lower limit less than the upper limit.
See section 3.3.

30

The partial boundary value exceeds the range of the span.

Set the boundary value in the range of “the minimum

span value + 1 digit” to “the maximum span value – 1

digit.”
See section 6.9

31

Partial-expansion display is set ON for a SKIPPED channel.

Cannot be specified on channels set to Skip.
See sections 3.3 and 6.9

35

The upper and lower limits of the display band are equal.

Set the upper limit greater than the lower limit + 5.
See section 6.6

36

The lower limit of the display band is greater than the upper

limit.

Set the upper limit greater than the lower limit + 5.
See section 6.6

37

The display band is narrower than 4% of the entire display.

Set the upper limit greater than the lower limit + 5.
See section 6.6

40

Incorrect group set character string.

Check the syntax.
See section 6.1

41

There is no specified input channel.

Specify a channel that is installed.
See sections 2.3 and 6.1

42

Exceeded the number of channels which can be set.

Up to 10 channels/group (MV2000).
Up to 6 channels/group (MV1000).
See section 6.1

43

A channel number cannot repeat in a group.

Check that a channel is not registered twice.
See section 6.1

45

There is no character string saved in the clipboard.

Copy a character string to the clipboard.
See section 6.1

Chapter 12 Troubleshootng and Mantenance

Advertising
This manual is related to the following products: