Yokogawa Button Operated MV2000 User Manual

Page 136

Advertising
background image

5-14

IM MV1000-01E

• INFORMATION (nformaton about the dsplayed measured data)

The following information is displayed.

Page switch mark

When using the batch function,

a comment field and a text field

are displayed on the second

and third pages.

Use the left and right arrow

keys to switch the page.

Item

Descrpton

Filename

Displays “Memory” for data stored to internal memory. Displays

the file name for data stored to an external storage medium.

File type

Displays “Display” for display data or “Event” for event data.

Serial number

The serial number of the MV that was used.

Batch number, Lot number Displayed for files that use batch functions.
Start time, End time

The time when recording started/stopped.

User name

The name of the user that operated the MV. This item is displayed

if the login function was used.

Note

If you display the measured data from an external storage medium, the serial number displayed
will be that of the MV that saved the data.

• Hstorcal Trend Background Color

You can change the background color of the historical trend display.

For configuration instructions, see section 6.12.

• TOP CHANNEL

The displayed scale marker and grid are those of the channel selected as the top

channel.

If you switch from TREND HISTORY display to another display, the TOP CHANNEL

setting is cleared. Channels other than the top channel are displayed in the order set
within their display groups. (Set in Group set, Trip line which can be opened from the
MENU tab after pressing MENU.)

5.3 Dsplayng Prevously Measured Data (TREND HISTORY)

Advertising
This manual is related to the following products: