Yokogawa Button Operated MV2000 User Manual

Page 13

Advertising
background image

1-4

IM MV1000-01E

The following input processing and math operations are available.

DC voltage

DC current

Shunt resistor

1 - 5 V

Burnout detection

Burnout detection

RJC

Square root computation

Thermocouple

RTD

Pulse
(/PM1 option)

ON/OFF input

Convert to number of pulses per second

Moving average

Linear

scaling

Calibration

correction

(/CC1 option)

Difference

computation

Low cut

Measured

value

Dotted lines indicate that the function can be selected.

Measured

value

Moving Average

External reference junction

Internal

reference

junction

Performs reference junction compensation on

a thermocouple.

Reference Junction Compensation

Burnout Detection

Detects and indicates a sensor burnout.

1 - 5 V

Detected on the

1-5 V value.

Burnout

For configuration instructions, see section 3.2.

For configuration instructions, see section 3.2.

For configuration instructions, see section 3.4.

Burnout

Indicated as

Burnout.

Indicated as

Burnout.

Thermoelectromotive

force

1

2

3

10.0 mV

5.0 mV

0.0 mV

5.0 mV

15.0 mV

10.0 mV

5.0 mV

10.0 mV

10.0 mV

15.0 mV

10.0 mV

11.7 mV

Clear

n

th

time

n + 1

th

time

n + 2

th

time

Clear

(Moving average)

New sampled

data

New sampled

data

Operation example when the number

of moving average data points is 3

Number of moving

average data points: 2 to 400

Sampling data

in the buffer

Sampling

Eliminates noise.

MV

Thermocouple

Thermocouple

1.2 Input