Yokogawa DA100 User Manual

Page 54

Advertising
background image

IM DA100-11E

6-2

6.1 Setting Command

Setting the strain input

Setting

SRp1, p2, p3, p4, p5<terminator>

p1 Channel number (001 to 560)
p2 STRAIN
p3 Measurement range

2k/20k/200k

p4 Left value of span
p5 Right value of span

Example

Set the measurement range of subunit 1 and
channel 01 to 20k, left value of span to 0, and
right value of span to 100.

SR101, STRAIN, 20k, 0, 100

Comments

• Even when the settings of p3 are the same

parameters, the measurement range of the strain
varies depending on the gauge method. For
details, refer to page 5-7.

• Set spans, p4 and p5, within the measurement

range indicated in page 5-7.

• Input values within 6 digits disregarding the

decimal point for p4 and p5.

• The measurement range cannot be changed

while executing a report or computation.

Setting the pulse input

Setting

SRp1, p2, p3, p4, p5, p6, p7, p8, p9<terminator>

p1 Channel number (001 to 560)
p2 PULSE
p3 Measuring mode

RATE momentary pulse count mode
GATE contact ON/OFF detection mode

p4 Left value of span (0 to 30000)
p5 Right value of span (0 to 30000)
p6 Left value of scale (0 to 30000)
p7 Right value of scale (0 to 30000)
p8 Decimal point position of scale (0 to 4)
p9 Filter (ON or OFF)

Example

Set subunit 1 and channel 01 to momentary pulse
count mode, left value of span to 0, and right value
of span to 1000.

SR101, PULSE, RATE, 0, 1000, 0, 1000, 0

Comments

• When omitting p6, p7, or p8, omit all three.
• When summing the count number or the ON

time, set the computational expression with the
SO command, then execute computation start
with the EX command. When the pulse input
module is installed, the use of computation
channels become available without the math
option. However, the only computational
expression available in this case is
TLOG.PSUM(XXX) for summing the count
number or the ON time.

• The measurement range cannot be changed

while executing a report or computation.

Setting Scaling

Setting

SRp1, p2, p3, p4, p5, p6, p7, p8, p9<terminator>

p1 Channel number (001 to 560)
p2 SCL
p3 Type of input

VOLT DC voltage
TC Thermocouple
RTD Resistance temperature detector
DI Contact

p4 Measurement range
p5 Left value of span
p6 Right value of span
p7 Left value of scale (-30000 to 30000)

p8 Right value of scale (-30000 to 30000)
p9 Decimal point position of scale (0 to 4)

Example

Change channel 02 of subunit 0 to an input in
which 1 V is converted to 0.00 and 5 V to 100.00.

SR002, SCL, VOLT, 6V, 1000, 5000, 0, 10000,
2

Comments

• For the p4 measurement range, see the range

parameter table on Pages 5-6 and 5-7.

• For the p5 and p6 setting spans, set them in the

ranges shown in the setting range on Pages 5-6
and 5-7.

• For p5 and p6 setting values, enter them within

6 digits excluding the decimal point.

• Select either to set all three parameters, p7, p8

and p9, or to omit them all.

• The measurement range cannot be changed

while executing a report or computation.

Setting the power monitor scaling

Setting

SRp1, p2, p3, p4, p5,...., p11<terminator>

p1 Channel number (001 to 560)
p2 SCL
p3 AC
p4 Wiring method

1Ph2W:

Single-phase two-wire system

1Ph3W:

Single-phase three-wire system
(only on a three-wire system)

3Ph3W-2I: Three-phase three-wire system

(only on a two-voltage two-
current three-wire system)

3Ph3W-3I: Three-phase three-wire system

(only on a three-voltage three-
current three-wire system)

3Ph4W:

three-phase four-wire system
(only on a three-wire system)

p5 Input range

25 V - 0.5 A
25 V - 5 A
250 V - 0.5 A
250 V - 5 A

p6 Measurement item
p7 Left value of span
p8 Right value of span
p9 Left value of scale (-30000 to 30000)
p10 Right value of scale (-30000 to 30000)
p11 Decimal point position of scale (0 to 4)

Example

Measure the active power on subunit 0 and channel
01, and convert 10W to 0.00 and 100W to 100.00.

SR001, SCL, AC, 3Ph3W-2I, 250V-0.5A,

P1, 10, 100, 000, 10000, 2

Comments

• Settings of p4 and p5 are common within one

power monitor module. If the setting is
changed, the settings on other channels also
change.

• For the measurement item p6, refer to the list

of measurement items on page 5-9.

• There are items that can not be set on the

measurement item, p6, depending on the wiring
method or the settings on other channels. Set
within the group indicated on page 5-9.

• Set the span values, p7 and p8, within the

measurement range indicated on page 5-7.

• Input values within 6 digits disregarding the

decimal point for p7 and p8. The decimal point
position is fixed to the decimal point position
of the setting range of page 5-7.

• For p9, p10, and p11, either set all the three

parameters or omit all three.