Cm sets communication input data, Xv sets the measurement period – Yokogawa DA100 User Manual

Page 58

Advertising
background image

IM DA100-11E

6-6

6.1 Setting Command

CM

Sets communication input data.

Mode

Operation mode

Setting

CMp1, p2 <terminator>

p1

Communication input data number:
Stand-alone model : C01 to C30
Expandable model : C01 to C60

P2

Numeric value: -32000 to 32000

Example

Set 300 to the communication input data number
C10.

CMC10, 300

Comments

• This command is effective only with the

optional computation functions.

• he position of the decimal point is selected

according to the position of the decimal point
set for the span with an SO command. So you
can set communications input data without
taking the decimal point into consideration.

RO

Setting hourly/daily/monthly report to
ON/OFF and the time to create the
report

Mode

Setup mode

Setting

ROp1, p2, p3, p4<terminator>

p1 ON Create hourly report

OFF Hourly report OFF

p2 ON1 Create daily report in standard format

ON2 Create daily report in extended format
OFF Daily report OFF

p3 ON1 Create monthly report in standard

format

ON2 Create monthly report in extended

format

OFF Monthly report OFF

p4 Date and time to create the report

DD HH

DD: Date
HH: Hour

Example

Do not create hourly reports and create daily and
monthly reports in standard format. Create the
monthly report at 10 O'clock on the first day of
every month and the daily report at 10 O'clock
everyday.

ROOFF, ON1, ON1, 01 10

Comments

• Effective only on instruments with the optional

report function.

• The report is output using TS4+device trigger

(GET)+RF.

• Date is set between 01 and 28 and hour is set

between 00 and 23 for p4.

RM

Setting report channel to ON/OFF and
the report computation type

Mode

Setup mode

Setting

RMp1, p2, p3, p4, p5<terminator>

p1 Report channel number (R01 to R60)
p2 Set the report channel to ON/OFF.
p3 Corresponding channel number

Measurement channel: 001 to 060 for stand-

alone type, 001 to 560 for extended type

Computation channel: A01 to A30 for

stand-alone type, A01 to A60 for extended
type

p4 The type of computation

INST

Instantaneous value at the time

when the report is created.

AVE

Average value

SUM

Summed value

p5 Conversion of the standard unit of time

INTVL No conversion

/sec Convert as the summed value of the

physical amount every second

/min Convert as the summed value of the

physical amount every minute

/hour Convert as the summed value of the

physical amount every hour

/day Convert as the summed value of the

physical amount everyday

Example

Set the summed value of the measurement channel
005 which is measuring the amount of flow in units
of l/min to the report channel, R02.

RMR02, ON, 005, SUM, /min

Comments

• Effective only on instruments with the optional

report function.

• If average value (AVE) is set on p4, minimum

and maximum values are also computed along
with the average value.

• If SUM is set on p4, the following computations

are carried out along with the summed value
from the time when the report was created to
the next report.
For hourly report: Total of the summed values
up to the time the daily report is created.
For daily report: Total of the summed values
up to the time the monthly report is created
Only summed value for monthly report

• The settings p3, p4 and p5 are effective when

p2 is ON.

• The setting p5 is effective only when p4 is set

to SUM.

• Since the sum is computed every measurement

interval using the sampled data, there are cases
when the physical amount per unit amount of
time is measured, the computed result differs
from the actual summed value (the
measurement interval and the unit time are
different). In this case, set the same unit used
for the unit time of the physical amount being
measured to p5. The summed values are
computed using the following converting
equations depending on the parameters.

INTVL

(measured values)

/sec

(measured values)

x

measurement

interval

/min

(measured values)

x

measurement

interval/60

/hour

(measured values)

x

measurement

interval/3600

/day

(measured values)

x

measurement

interval/86400

The unit of the measurement interval is seconds.

XV

Sets the measurement period.

Mode

Setup mode

Setting

XVp1<terminator>

p1

Measurement interval (0.5, 1, 2, 3, 4, 5,
6, 10, 12, 15, 20, 30, or 60)

Example

Set the measurement period to 10 seconds.

XV10

Comments

• For the measurement period setting time, the

parameter that can be set differs depending on
the A/D integration time and the filter on/off
setting. For each condition, the longest
measurement period is the upper limit. The
longest periods are shown below.