Output the measurement data (binary code) – Yokogawa DA100 User Manual

Page 92

Advertising
background image

IM DA100-11E

8-11

Sample Program

8

Output the Measurement Data (Binary Code)

Read out the measurement data by BINARY code from DA100, display on CRT of the personal
computer, and save to floppy disk.

10

‘

20

OPEN “COM1:E81N” AS #1

30

OPEN “TS0BIN.DAT” FOR OUTPUT AS #2

40

‘

50

PRINT #1,CHR$(&H1B)+”O 01"

60

LINE INPUT #1,D$ :PRINT D$

70

PRINT #1,”TS0"

80

LINE INPUT #1,D$ :PRINT D$

90

PRINT #1,”BO1"

100

LINE INPUT #1,D$ :PRINT D$

110

PRINT #1,CHR$(&H1B)+”T”

120

LINE INPUT #1,D$ :PRINT D$

130

PRINT #1,”FM1,001,010"

140

D$=INPUT$(2,#1)

150

PRINT #2,D$

160

A=CVI(MID$(D$,1,2))

170

PRINT A

180

D$=INPUT$(A,#1)

190

PRINT #2,D$

200

PRINT ASC(MID$(D$,1,1)); :PRINT “/”;

210

PRINT ASC(MID$(D$,2,1)); :PRINT “/”;

220

PRINT ASC(MID$(D$,3,1)); :PRINT

230

PRINT ASC(MID$(D$,4,1)); :PRINT “:”;

240

PRINT ASC(MID$(D$,5,1)); :PRINT “:”;

250

PRINT ASC(MID$(D$,6,1))

260

‘

270

L=0

280

FOR I=7 TO A

290

PRINT RIGHT$(“0”+HEX$(ASC(MID$(D$,I,1))),2)+” “;

300

L=L+1

310

IF L=5 THEN L=0 : PRINT

320

NEXT I

330

‘

340

PRINT #1,CHR$(&H1B)+”C 01"

350

LINE INPUT #1,D$ :PRINT D$

360

CLOSE

370

END

8.3 RS-422-A/RS-485 Sample Programs

Advertising