Yokogawa DA100 User Manual

Page 67

Advertising
background image

IM DA100-11E

7-2

7.2

Measured/Computed Data Output Format (ASCII
code)

The data are output in the following format by receiving TS0 + “Device Trigger (GET)” + FM0/
FM2:

DATEYYMMDDCrLf
TIMEhhmmssCrLf
S1S2A1A1A2A2A3A3A4A4UUUUUUCCC,

±

DDDDDE - ECrLf

Each symbol denotes the following:

YY:

Year

MM:

Month

DD:

Day

hh:

Hour

mm:

Minute

ss:

Second

S1:

Data status 1

E ------------- Abnormal
N ------------ Normal
D ------------ Differential input
O ------------ Over
S ------------- Skip or computation channel off

S2:

Data status 2

Space ------- Interim data
E ------------- Last data

A1A1:

Alarm status (level 1)

A2A2:

Alarm status (level 2)

A3A3:

Alarm status (level 3)

A4A4:

Alarm status (level 4)

H

--------- Upper-limit alarm

L

---------- Lower-limit alarm

dH ----------- Upper-differential-limit alarm
dL ----------- Lower-differential-limit alarm
RH ---------- Increasing rate-of-change limit alarm
RL ----------- Decreasing rate-of-change limit alarm

UUUUUU: Unit

mV

--- mV

V

-- V

C

--

°

C

UUUUUU ------- Arbitrary

CCC:

Channel number

±

:

Data polarity (+, -)

DDDDD:

Data mantissa (8 characters for the computation chanel

±

99999 ----- Over data

+99999 ----- Abnormal data

E - E:

Data exponent

Note

• Data in the channel not connected in the system settings, including channel numbers, are not output.

• In a channel for data computation, the channel number CCC is indicated as follows:

Stand-alone model: A01 to A30

Expandable model: A01 to A60