Yokogawa DA100 User Manual

Page 84

Advertising
background image

IM DA100-11E

8-3

Sample Program

8

Output the measurement data (BINARY Code)

Read out the measurement data by BINARY code from DA100, display on CRT of personal computer,
and save to floppy disc.

10

'TS0 BO1 <GET> FM1

20

OPEN "TS0BIN.DAT" FOR OUTPUT AS #1

30

ISET IFC

40

CMD DELIM=0

50

PRINT @1;"TS0"

60

PRINT @1;"BO1"

70

WBYTE &H3F,&H21,&H8,&H3F;

80

PRINT @1;"FM1,001,010"

90

CMD DELIM=3

100

LINE INPUT @1;D$:PRINT #1,D$

110

A=CVI(MID$(D$,1,2)):PRINT A

120

L=0

130

PRINT ASC(MID$(D$,3,1));:PRINT "/";

140

PRINT ASC(MID$(D$,4,1));:PRINT "/";

150

PRINT ASC(MID$(D$,5,1));:PRINT

160

PRINT ASC(MID$(D$,6,1));:PRINT ":";

170

PRINT ASC(MID$(D$,7,1));:PRINT ":";

180

PRINT ASC(MID$(D$,8,1));:PRINT

190

L=0

200

FOR I=6 TO A-1

210

PRINT RIGHT$("0"+HEX$(ASC(MID$(D$,I+3,1))),2)+" ";

220

L=L+1

230

IF L=6 THEN L=0 :PRINT

240

NEXT I

250

CLOSE:STOP

260

END

Output the system configuration data

Read out the configuration data from DA100, display on CRT of personal computer, and save to
floppy disc.

10

'TS5 <GET> CF

20

OPEN "TS5.DAT" FOR OUTPUT AS #1

30

ISET IFC

40

CMD DELIM=0

50

PRINT @1;"TS5"

60

WBYTE &H3F,&H21,&H8,&H3F;

70

PRINT @1;"CF0"

80

LINE INPUT @1;D$:PRINT D$:PRINT #1,D$

90

LINE INPUT @;D$:PRINT D$:PRINT #1,D$

100

IF LEFT$(D$,2)<>"E:" GOTO 90

110

CLOSE:STOP

120

END

8.1 GP-IB Sample Programs