Xt sets the temperature unit, Xz performs a/d calibration of the input module – Yokogawa DA100 User Manual

Page 61

Advertising
background image

IM DA100-11E

6-9

Commands

6

E0Cr+Lf: The processing of a received

command has completed normally.

E1Cr+Lf: There is an error in the received

command.

• After executing the command, the clock is

initialized to 96/01/01, 00:00:00.

XT

Sets the temperature unit

Mode

Setup mode

Setting

XTp1<terminator>

p1

Temperature unit
C

°

C

F

°

F

Example

Set temperature unit to

°

C

XTC

XZ

Performs A/D calibration of the input
module.

Mode

A/D calibration mode

Calibration using the actual input
value at the input terminal

Setting

XZp1, p2, p3, p4<terminator>

p1

Subunit number (0 to 5)

p2

Slot number (0 to 5)

p3

CAL/EXEC

p4

Calibration item
• For universal input module

20 mV, 60 mV, 200 mV, 2 V, 6 V, 20
V, 50 V, Pt:1 mA, Pt:2 mA, Pt:1 mA-
H, Pt:2 mA-H, Cu:2 mA

• For mA input module

20 mA

• For strain input module

2k_ZERO, 2k_SPAN, 20k_ZERO,
20k_SPAN, 200k_ZERO, 200k_SPAN

• For digital input module

60 mV, 6 V

Example

Execute a 20-mV range calibration on the module
in slot 1 of subunit 0.

XZ0, 1, CAL/EXEC, 20mV

Correcting the calibration value
manually (universal input module)

Setting

XZp1, p2.....p5<terminator>

p1

Unit number (0 to 5)

p2

Slot number (0 to 5)

p3

DISPLAY

p4

Zero calibration value for 20-mV range

p5

Span calibration value for 20-mV range

p6

Zero calibration value for 60-mV range

p7

Span calibration value for 60-mV range

p8

Zero calibration value for 200-mV range

p9

Span calibration value for 200-mV range

p10 Zero calibration value for 2-V range
p11 Span calibration value for 2-V range
p12 Zero calibration value for 6-V range
p13 Span calibration value for 6-V range
p14 Zero calibration value for 20-V range
p15 Span calibration value for 20-V range
p16 Zero calibration value for 50-V range
p17 Span calibration value for 50-V range
p18 Zero calibration value for Pt:1 mA
p19 Span calibration value for Pt:1 mA
p20 Zero calibration value for Pt:2 mA
p21 Span calibration value for Pt:2 mA
p22 Span calibration value for Pt:1 mA-H

p23 Span calibration value for Pt:2 mA-H
p24 Zero calibration value for Cu:2 mA
p25 Span calibration value for Cu:2 mA

Correcting the calibration value
manually (DCV/TC/DI input module)

Setting

XZp1, p2.....p17<terminator>

p1

Unit number (0 to 5)

p2

Slot number (0 to 5)

p3

DISPLAY

p4

Zero calibration value for 20-mV range

p5

Span calibration value for 20-mV range

p6

Zero calibration value for 60-mV range

p7

Span calibration value for 60-mV range

p8

Zero calibration value for 200-mV range

p9

Span calibration value for 200-mV range

p10 Zero calibration value for 2-V range
p11 Span calibration value for 2-V range
p12 Zero calibration value for 6-V range
p13 Span calibration value for 6-V range
p14 Zero calibration value for 20-V range
p15 Span calibration value for 20-V range
p16 Zero calibration value for 50-V range
p17 Span calibration value for 50-V range

Correcting the calibration value
manually (mA input module)

Setting

XZp1, p2, p3, p4, p5<terminator>

p1 Unit number (0 to 5)
p2 Slot number (0 to 5)
p3 DISPLAY
p4 Zero calibration value for 20-mA range
p5 Span calibration value for 20-mA range

Correcting the calibration value
manually (Strain input module)

Setting

XZp1, p2, p3, ....., p9<terminator>

p1 Unit number (0 to 5)
p2 Slot number (0 to 5)
p3 DISPLAY
p4 Zero calibration value for 2k range
p5 Span calibration value for 2k range
p6 Zero calibration value for 20k range
p7 Span calibration value for 20k range
p8 Zero calibration value for 200k range
p9 Span calibration value for 200k range

Correcting the calibration value
manually (Digital input module)

Setting

XZp1, p2.....p7<terminator>

p1

Unit number (0 to 5)

p2

Slot number (0 to 5)

p3

DISPLAY

p4

Zero calibration value for 60-mV range

p5

Span calibration value for 60-mV range

p6

Zero calibration value for 6-V range

p7

Span calibration value for 6-V range

Example

Modify the span calibration value for the 6V range
of the module in slot 1 of subunit 0 to 32000.

XZ0, 1, DISPLAY, , , , 32000

Storage of A/D calibration data in a
module (A/D calibration END
processing)

Setting

XZp1, p2, p3, p4<terminator>

p1

Unit number (0 to 5)

p2

Slot number (0 to 5)

p3

END

6.1 Setting Command