Yokogawa PC-Based MX100 User Manual

Page 1220

Advertising
background image

26-2

IM MX190-01E

Error number Message string

Description

Corrective Action/Remarks

10

Not acknowledge

Received an unsupported response.

The response from the measurement instrument is different from the expected

response. Check the transmitted command or the procedure.

11

Not support

Specified an unsupported function.

May have specified a value outside the range. Check the values of the

parameters passed to the function.

12

Not data

There is no data.

Invalid input to the function. The data specified in the parameters may be

incorrect. Or, if the parameter is a string, the string may not be correct. Check

the values of the parameters passed to the function.

13

Exception

An exception occurred.

A system exception may have occurred, or field may have failed to be

established. Check the PC environment.

26.1 API Error Messages