5.1.5. Irradiance
1. Move the cursor to “Irradiance” by means of the arrow

keys (

,) and confirm with ENTER.

2. The display shows the screen with the options: “ Irr min

IV which allows to set the minimum threshold of
measured irradiance expressed in W/m

2

, used by meter as

reference during the I-V curve measurements and “Irr min
EFF
” (only SOLAR I-V) which allows to set the minimum
threshold of measured irradiance expressed in W/m

2

, used

by meter as reference during the check PV efficiency
measurements

3. Set the value by means of the arrow keys (

,). The

accuracy indicated in this manual is granted under
condition indicated in § 10.1 and § 10.2. In I-V curve
measurement the value is set within the 0

800 W/m

2

interval while 400

800 W/m

2

for the PV check

operations (only SOLAR I-V)

4. Confirm with SAVE and the “Saved data” message is

displayed for a while. Press ESC/MENU key to exit without
saving and to go back to the previous screen.

15/05/10 15:34:26

I r r m i n I V

:  0  W / m 2

I r r m i n E F F

:  6 0 0  W / m 2

SAVE to store data

S E T


CAUTION

The “0 W/m2” setting for parameter “Min Irr IV” allows carrying out I-V
measurements without the following conditions being checked:
 Connection of reference cell to input IRR of the instrument
 Unsteady irradiation values
 Number of modules consistent with the measured open circuit voltage



5.1.6. DC Clamp (only SOLAR I-V)
This option allows setting a possible correction factor K for the DC clamp, in order to improve
current measurement. If present, the correction factor is indicated on the rear label of the clamp
itself, indicated as:

K= X.xxx

In case no label is present, set k = 1.000

1. Move the cursor to “DC Clamp” by means of the arrow

keys (

,) and confirm with ENTER

2. The item: “DC clamp k” is shown at display to set the

corrective factor in a interval within 0.950 and 1.050. Set
the value by means of the arrow keys (

, )

3. Confirm with SAVE and the “Saved data” message is

displayed for a while. Press ESC/MENU key to exit without
saving and to go back to the previous screen

15/05/10 15:34:26

D C c l a m p k

:  1 . 0 0 0 

SAVE to store data

S E T