10.2. I-V TEST TECHNICAL FEATURES (I-V AND IVCK)

I-V: DC Voltage @ OPC

Range [V] (***)

Resolution [V]

Uncertainty (***)

5.0

 999.9

0.1

(1.0rdg + 2dgt)

(***) The I-V curve and Rs measurements start for VDC > 15V and the accuracy is defined for VDC > 20V

I-V: DC Current @ OPC

Range [A]

Resolution [A]

Uncertainty

0.10

 10.00

0.01

(1.0rdg + 2dgt)

I-V: DC Power @ OPC (Vmpp > 30V, Impp > 2A)

Range [W] (*, **)

Resolution [W]

Uncertainty

50

 9999

1

(1.0rdg + 6dgt)

Vmpp = Maximum power voltage, Impp = Maximum Power Current
(*) Max measurable value of Power must include FF value(~ 0.7)

 Pmax = 1000V x 10A x 0.7 = 7000W

(**) Test is stopped and the message “Thermal instability” occurs if the instrument detects Voltage > 700V and Current I >3A, I > -
0.038*V + 37.24 - 0.5

I-V: DC Voltage @ STC, IVCK: DC Voltage @ OPC and @ STC

Range [V]

Resolution [V]

Uncertainty (*,**)

5.0

 999.9

0.1

4.0rdg + 2dgt)

I-V: DC Current @ STC, IVCK: DC Current @ OPC and @ STC

Range [A]

Resolution [A]

Uncertainty (**)

0.10

 10.00

0.01

(4.0 rdg + 2 dgt)

I-V: DC Power @ STC (Vmpp > 30V, Impp > 2A)

Range [W]

Resolution [W]

Global Uncertainty (**)

50

 9999

1

(5.0 rdg + 1 dgt)

Vmpp = Maximum power voltage, Impp = Maximum Power Current
(*) Measurements start for VDC > 15V and the accuracy is defined for VDC > 20V
(**) Test conditions:
 Test cond.: Steady Irrad.

700W/m

2

, spectrum AM 1.5,solar incidence vs perpendicular.

  25°, Cells Temp. [15..65°C]

 Global accuracy include contribute of solar sensor and its measuring circuit

Irradiance (with reference cell)

Range [mV]

Resolution [mV]

Accuracy

1.0

 100.0

0.1

(1.0rdg + 5dgt)

Temperature (with auxiliary probe)

Range [°C]

Resolution [°C]

Accuracy

-20.0

 100.0

0.1

(1.0rdg + 1°C)


10.3. SAFETY SPECIFICATIONS
10.3.1. General
Instrument safety:

IEC/EN61010-1

Technical literature:

IEC/EN61187

Accessory safety:

IEC / EN61010-031

Measurements:

IEC/EN60891 (I-V curve test)

IEC/EN 60904-5 (Temperature measurement)

Insulation: double

insulation

Pollution level:

2

Category of measure:

CAT II 1000V DC, CAT III 300V to earth

Max 1000V among inputs P1, P2, C1, C2