Caution – HT instruments SOLAR I-V User Manual

Page 54

Advertising
background image

I-V400 - SOLAR I-V

EN - 52

6.3.3. Fast check IVCK without irradiance measurement


CAUTION

 The maximum voltage among inputs P1, P2, C1 and C2 is 1000V DC. Do not

measure voltages exceeding the limits expressed in this manual.

 The maximum current allowable for the instrument is 10A. Do not effect tests

in parallel on FV module strings.

 Never effect tests on modules or FV strings connected to DC/AC converter.

1. Turn on the instrument pressing ON/OFF
2. Check that remote unit SOLAR-02 is not selected (see § 5.1.4– setting NO)
3. Check that the value of minimum irradiance set in the section “Irradiance” (see §5.1.5)

is equal to 0

4. Go back to main MENU and select “I-V
5. Select “I-V” and press ENTER to enter the section of FV module performance

measurement. Generally when selecting I-V on general menu the instrument automatically
shows the last mode under use (I-V or IVCK).

6. In case the current mode is not IVCK, press ENTER,

select “Measurement ” by means of arrow keys
(

,). Use the arrow key  to enter the internal

submenu, select “IV Check” and confirm pressing
ENTER to start up the initial screen of fast check IVCK.

15/05/10 15:34:26

V d c = 0 . 0 V

I r r = 0 W / m 2

T c = - - - ° C

Module: SUNPOWER 210

F e a t u r e

I V

I m p

M e a s u r e m e n

t R S

M i s

I V C h e c k

Selection

I - V

7. Check the preliminary settings based on descriptions of § 6.3.2
8. Connect the instrument to the module/string under test as shown in the following

picture. In particular connect the negative pole of the module/string to terminals P1, C1
and it positive pole to terminals P2, C2.

CAPTION:

P1: Black cable
P2: Blue cable
C1: Green cable
C2: Red cable

1:

Module or FV string

Fig. 10: Connection to module/FV string for IVCK tests without irradiance measurement