Caution – HT instruments SOLAR I-V User Manual

Page 35

Advertising
background image

I-V400 - SOLAR I-V

EN - 33

6.2. I-V CURVE MEASUREMENT
The meter can perform I-V curve tests by using one of the herewith modes:

 I-V curve test with Irr/Temp measurement directly performed by the instrument
 I-V curve test with Irr/Temp measurement performed by use of remote unit SOLAR-02

For theoretical information on measurement see the § 11.1

6.2.1. I-V curve measurement by use of instrument

CAUTION

 The maximum voltage among P1, P2, C1, and C2 inputs is 1000V DC. Do

not measure voltages exceeding the limits prescribed by this manual

 The maximum current measured by the meter is 10A. Do not perform test

on strings of PV modules in parallel with them

 Do not perform test on PV modules or strings connected to DC/AC

converter

1. Turn on the meter by pressing ON/OFF key
2. Check that option NO relative to the selection of remote unit SOLAR-02 should be set

on the instrument (see § 5.1.4)

3. Check that the values set on the “Remote unit” section (see § 5.1.4) should be

consistent with the reference cell depending on the kind of module/string on test

4. Check that the value set on “Irradiance” section (see § 5.1.5) should be consistent with

the measurement to perform. It is suggested to perform the measurement with a
threshold

 700 W/m

2

in compliance to the IEC/EN60891 guideline

5. Back to the main menu and select the “I-V” option

6. Press ENTER, by means of arrow keys (

,) select

item “Meas. Type ”

7. Using the

 key, select item “I-V Test” and confirm by

pressing ENTER

15/05/10 15:34:26

V d c = 0 . 0 V

I r r = 0 W / m 2

T c = - - - ° C

Module: SUNPOWER 210

I - V T e s t

S e t R S

T e s t

M e a

I V F a s t C h e c k

Select

I - V

8. The instrument displays a screen similar to the one

reported here to the side, where:
 Vdc = DC output voltage from module measured

between C1 and C2 inputs of meter

 Irr = irradiance measured by supplied reference cell
 Tc = temperature of module cell. The herewith options

are available for this field:

 AUTO  automatic mode of temperature measurement
 Number  MAN or AUX mode of temp. measurement
 “ - - - “  AUX mode with no connected probe
 Module = type of selected module
 Temp = measurement mode of module temperature

15/05/10 15:34:26

V d c = 0 . 0 V

I r r = - - - W / m 2

T c = A u t o

M o d u l e : P A N E L 0 1

Temp: Auto

Select

I - V

Advertising
This manual is related to the following products: