18.

On SOLAR I-V, press key (

) to access the second

screen which contains the values of the output DC
parameters of the strings according to the number of DC
inputs set (see § 5.2.1.1). In particular, in this screen:

 Vdcx = DC voltage of string x
 Idcx = DC current of string x
 Pdx = DC power of string x.

We recommend checking that the values of the electrical
parameters (Vdc, Idc, Pdc) are consistent with the system
being measured.

15/05/10 15:34:26

V d c 1

4 6 0 . 1

k W

V d c 2 4 6 1 . 4

V

V d c 3

4 6 2 . 5

A

I d c 1

2 . 2 5

A

I d c 2

2 . 3 1

A

I d c 3

2 . 2 1

A

P d c 1

1 . 0 3 5

k W

P d c 2

1 . 0 6 6

k W

P d c 3

1 . 0 2 4

k W

G O – s t a r t r e c

Select

M P P

19. On SOLAR I-V, press key (

) to access the third screen

which contains the values of the electrical parameters on
the AC side of the inverter, consistently with the settings
made in § 5.2.2 (single-phase, three-phase 4 wires). In
particular, in this screen:

 Vacxy = AC voltage between Phase and Neutral (if

single-phase) or between Phases x and y (if three-
phase)

 Iacx = AC current of phase x
 Pacx = AC power of phase x

We recommend checking that the values of the electrical
parameters (Vac, Iac, Pac) are consistent with the system
being measured.

15/05/10 15:34:26

V a c 1 2 4 0 1 . 4

V

V a c 2 3 4 0 1 . 1

V

V a c 3 1 4 0 0 . 1

V

I a c 1

4 . 2 6

A

I a c 2

4 . 2 6

A

I a c 3

4 . 2 7

A

P a c 1

9 8 7

W

P a c 2

9 8 6

W

P a c 3

9 8 5

W

G O – s t a r t r e c

Select

M P P

E x a m p l e o f a s c r e e n f o r

P V s y s t e m s w i t h t h r e e -

p h a s e o u t p u t

20.

Keeping the three instruments near each other (max
distance approx. 1m), press the GO/STOP key on SOLAR I-
V to start testing. Consequently:

 The display of SOLAR I-V shows the message “rec. start

waiting

 The display of SOLAR-02 shows the message “HOLD

and the time, expressed in seconds, remaining before the
recording is started

 On MPP300, the STATUS LED turns on green (not

flashing)

15/05/10 15:34:26

P R p

- - -

I r r

- - -

W / m 2

P n o m 3 . 5 0 0

k W

T c

- - -

° C

T e

- - -

° C

P d c

3 . 1 2 5

k W

P a c

2 . 9 6 0

k W

n d c

- - -

n a c

0 . 9 5

R e c . r u n n i n g

Select

M P P

21.

Upon reaching the instant “00” after pressing the
GO/STOP key, the test is started and the three units are
synchronized with each other. In these conditions:

 The display of SOLAR I-V shows the message “rec.

running

 The display of SOLAR-02 shows the message

Recording…

 On MPP300, the STATUS LED flashes green

15/05/10 15:35:00

P R p

- - -

I r r

- - -

W / m 2

P n o m 3 . 5 0 0

k W

T c

- - -

° C

T e

- - -

° C

P d c

3 . 1 2 5

k W

P a c

2 . 9 6 0

k W

n d c

- - -

n a c

0 . 9 5

R e c . r u n n i n g

Select

M P P